Abstract
Early detection of primary and metastatic cancer has a profound influence on prognosis for all cancers. Currently, imaging CT/MRI and/or 18F-FDG-PET followed by tissue biopsy is the gold standard for detection in clinical practice, depending on the lesion site. Specificity and sensitivity of imaging modalities, depending on the type of cancer and its site, remain challenging. Here, we report on a developing technology, magnetic particle imaging (MPI), for the detection of primary tumors and metastases using a mouse model of breast cancer. Using this technique, we identified multiple metastatic lesions in mice bearing luciferase-expressing 4T1 breast tumors, which were not clearly defined by bioluminescence or microCT imaging. Post-processing of imaging data using an improved algorithm yielded higher-resolution images, enabling quantification of signal intensity of micrometastatic lesions throughout the mouse body.
Key features
• Commercial SynomagTM nanoparticles coated with poly(ethylene glycol) by the manufacturer were used to detect mammary tumors and metastases in a mouse model of breast cancer.
• Magnetic particle imaging, a new tracer imaging technique, was used for detecting cancer lesions.
• With improved processing algorithms, we were able to detect microscopic lesions that could not be seen with micro-computed tomography or bioluminescent imaging.
• MPI promises to provide another imaging modality for cancer.
Keywords: Magnetic particle imaging (MPI)Graphical overview
Iron oxide nanoparticles accumulating in the primary and metastatic tumor microenvironment enable their non-invasive detection by magnetic particle imaging
Background
Magnetic particle imaging (MPI) is a new tracer imaging technique that generates signals from magnetic materials (tracers) when exposed to an oscillating magnetic field [1]. Unlike computed tomography scans, which use X-rays, a sample scanned with MPI is not exposed to ionizing radiation, making it more suitable for living subjects. Superparamagnetic iron oxide (SPIO) nanoparticles are shown to be ideal tracers for generating quality MPI signals [2,3]. To create an image, the MPI scanner uses opposing magnets to generate a field-free region (FFR), which is then rastered over the sample [2]. The scanner then uses a low frequency/low amplitude alternating magnetic field to force the tracer’s magnetic moments within the FFR to oscillate and generate a potential that is proportional to the number of nanoparticles. Outside the FFR region, SPIOs are magnetically saturated and produce no signal. The receiver coil detects the voltage created by the SPIOs when their magnetic moments oscillate 180° from their current voxel due to the superimposed alternating magnetic field when the FFR is scanned across their location. This signal, picked up by the receiver coil, can then be reconstructed as an image using reconstruction algorithms like harmonic-space or x-space [2,4]. This method exhibits high sensitivity, enabling the quantitative detection of nanogram tracer quantities. It is also unaffected by air or other void volumes, making it suitable for imaging lung metastases; locations devoid of the nanoparticle tracer generate no signal, reducing or eliminating background noise. MPI also does not generate signal from tissue (which is weakly diamagnetic) and thus must be co-registered with computed tomography (CT) or magnetic resonance imaging (MRI).
Here, we describe methods to detect tumors and metastases as reported in our previous study [5]. Targeting, or directing, nanoparticles to tumors has historically presented a challenge. Our recent research has shown that SynomagTM nanoparticles, coated with poly(ethylene glycol), are taken up by inflammatory cells and tissues (e.g., macrophages, fibroblasts, and collagen) in the tumor microenvironment, thereby enabling tumor detection by MPI [5]. We recently showed that SynomagTM nanoparticles, which are not superparamagnetic, are instead aggregated multi-core ferrite nanoparticles, or nanoflowers with magnetic anisotropy that makes them suitable tracers for MPI [6].
More recently, we reported that antibody-conjugated nanoparticles also accumulate in tumors and other non-cancer sites of inflammation, apart from organs of the reticuloendothelial system (liver and spleen). These non-cancer sites can be distinguished from tumors because they have a uniform and bright signal, whereas cancer tumors show a distinct outer ring pattern that distinguishes them [7]. This unique pattern to distinguish cancer from non-cancer tissues was used to detect tumors. In all cases, cancer was confirmed with post-mortem histopathology.
Materials and reagents
Biological materials
1. BALB/c mice (8-week-old female BALB/c mice from Jackson Labs, stock no: 000651) with primary tumor and metastases
2. Breast cancer cells -4T1-luc (gift from Dr. Saraswati Sukumar, Johns Hopkins University); alternatively, cells can be purchased from American Type Culture Collection (ATCC) 4T1-Luc2, catalog number: CRL-2539-LUC2
Reagents
1. Hydroxyethyl starch- and poly(ethylene glycol)-coated iron oxide nanoparticles, Synomag®-PEG (SPEG) nanoparticles (Micromod Partikeltechnologie, catalog number: 105-124-102); store at 4 °C
Note: SPEG nanoparticles are composed of 50-nm aggregated multi-core ferrite nanoparticles or nanoflowers with iron oxide ferrite cores coated with hydroxyethyl starch as the shell material, resulting in a hydrodynamic diameter of 100 nm. This was then covalently bonded to polyethylene glycol (PEG 25000-OMe) [5,6].
2. VivoTrax nanoparticles (Magnetic Insight, catalog number: MIVT01-LOT00009); store at 4 °C
Note: VivoTrax is a preclinical superparamagnetic iron oxide nanoparticle with a magnetite iron core and dextran coating similar to Ferucarbotran/Resovist, a clinically approved liver MRI contrast agent. The core iron size of VivoTrax is 4.2 nm, and the hydrodynamic diameter is 62 nm [7]. In vitro MPI signal and resolution characteristics of SPEG and VivoTrax showed that SPEG is superior to VivoTrax [5].
3. Roswell Park Memorial Institute (RPMI 1640) medium (Thermo Fisher Scientific, catalog number: 11875093)
4. Fetal bovine serum (FBS) (MilliporeSigma, catalog number: TMS-013-B)
5. Trypsin EDTA 1×, 0.25% (Thermo Fisher Scientific, catalog number: 25200056)
6. PierceTM D-Luciferin, monosodium salt, 1 g (Thermo Fisher Scientific, catalog number: 88292)
7. Isoflurane (MWI Animal Health, catalog number: NDC 13985-528-40)
8. Phosphate-buffered saline (PBS) (Quality Biologicals, catalog number:114-058-101CS)
Laboratory supplies
1. BioLite cell culture-treated flasks (T75) (Thermo Fisher Scientific, catalog number: 130190)
2. Corning® Costar® Stripette® serological pipettes, individually paper/plastic wrapped, capacity 5 mL (MilliporeSigma, catalog number: CLS4487)
3. Tips LTS 20 μL filter 960/5 SR-L10F (Rainin, catalog number: 17005860)
4. NuncTM conical sterile polypropylene centrifuge tubes (racked) (Thermo Fisher Scientific, catalog number: 339651)
5. 5 mL tubes (Fisher Scientific, catalog number: 50-302-8556)
6. Exel, 1cc TB syringe w/ 25g × 5/8" Needle (Exelint International, catalog number: EXE26044)
7. BD PrecisionGlideTM needle, 30 G × 1/2 in. (Becton Dickinson, catalog number: 305106)
8. FisherbrandTM sterile alcohol prep pads (Fisher Scientific, catalog number: 22-363-750)
9. Small tubes, MicroAmp fast reaction tube with cap (Applied Biosystems, catalog number: 4358297)
10. Eye lube (GenTeal tears, Alcon Inc.)
11. Paper surgical tape, 3" × 10 yds (Thomas Scientific, catalog number: 1152J14)
12. Ice
13. Tube holders and racks
Equipment
1. Cell culture incubator ((Thermo Fisher Scientific, model: HeracellTM VIOS 160i CO2 Incubator)
2. Centrifuge (Eppendorf, model: 5424R)
3. Cell counter (Denovix, model: S-10696)
4. IVIS Imaging system (PerkinElmer IVIS In Vivo Imaging System, model: 124262)
5. Tailveiner® Restrainer for Mice (Braintree Scientific, model: TV-150)
6. Isoflurane vaporizer (Midmark, model: Matrx VIP 3000-91305418)
7. Small Animal Monitoring & Gating System (SA Instruments, model: 1025T-FOT)
8. MOMENTUM MPI scanner (Magnetic Insight)
9. IVIS Spectrum CT System (Perkin-Elmer)
Software and datasets
1. Living ImageTM (Revvity, USA)
2. MATLAB (The MathWorks, Inc., Natick, MA, USA); a recent MATLAB version with GPU computing support is required for the reconstruction code used in this protocol
3. Momentum scanner software (Magnetic Insight, Inc.), installed on the computer supplied with the Momentum MPI scanner
4. MagImage (Magnetic Insight, Inc.), 3D Slicer, or equivalent image analysis software may be used for visualization, image registration, segmentation, and quantitation
5. MATLAB code for model-based 2D and 3D field-free-line MPI reconstruction is publicly available at https://github.com/toby-sanders/MPI-FFL-reconstruction
6. Experimental data associated with the original study are publicly available through the Johns Hopkins Research Data Repository: https://doi.org/10.7281/T1QZ3N5L
Procedure
A. Development of metastatic model
1. Grow 4T1-luc cells in RPMI media with 10% FBS in a T75 cell culture dish, incubating at 37 °C with 5% CO2.
2. On the day of injection, wash the cells with PBS and add 1 mL of trypsin to detach the cells.
3. Add 9 mL of RPMI with 10% FBS media and collect cells in a 50 mL centrifuge tube.
4. Spin down at 200× g for 2 min.
5. Discard the media and wash cells with PBS.
6. Count the number of cells using a cell counter.
7. Adjust the cell number to 0.025 × 106 in 50 μL of PBS.
8. Inject the cells orthotopically into the fourth mammary gland of mice under isoflurane anesthesia.
9. Monitor the tumor and metastatic growth by bioluminescence imaging weekly.
10. For this, inject mice intraperitoneally with freshly prepared D-Luciferin (150 mg/kg body weight) in PBS.
11. Wait 10 min and image mice after a 2-min exposure period in an IVIS imaging system under anesthesia.
12. Mice with confirmed tumor and metastasis are chosen for MPI.
B. Nanoparticle injection
1. Inject 30 mg/kg body weight of SPEG in 100 μL of sterile PBS (or desired concentration).
2. Restrain the mouse bearing 4T1-luc tumors/metastases in a mouse tail vein restrainer for injection.
2. Inject 100 μL of the solution intravenously through the tail vein into the mouse using a 30 G needle on a 1 mL syringe.
3. Wait for 72 h for near-complete clearance of nanoparticles (NPs) from the bloodstream. Alternatively, this duration can be chosen from prior blood-clearance experiments and may vary by nanoparticle type or coating.
C. In vivo magnetic particle imaging setup
Note: Please refer to the Momentum User Manual and Advanced User Manual for detailed operation of the MPI.
1. Prepare the mouse bed for holding the mouse during the scan. The setup consists of i) a mouse holder, ii) two anesthesia lines for incoming gas and outgoing gas to the collection canister, iii) two waterline connections that run circulating water at a temperature of 60 °C to keep the mouse warm, iv) a fiber optic temperature probe, and v) a respiration monitoring system (Figure 1). Once set up, turn on the circulating water chiller to begin heating the mouse holder.

Figure 1. Mouse preparation for magnetic particle imaging (MPI) scans. (A) Momentum scanner stage showing the whole setup of the mouse holder. (B–D) Anesthesia area and chamber. (E) Circulating water bath set at 60 °C. (F, G) Rectal temperature and respiration monitor setups.
2. Prepare imaging fiducials for later co-registration. Each fiducial consists of a ~1 μL drop of stock Vivotrax (5.5 mg/mL) nanoparticles in a small polystyrene tube (Figure 2a–c). A total of five fiducials should be used for each scan and taped to the holder in a “C” shape to aid with co-registration.
3. Switch on the main MPI breaker, press the power button twice to initiate the power-up sequence, and turn on the chiller.
4. As the MPI initializes, anesthetize the mouse using the isoflurane vaporizer.
5. Take out the anesthetized mouse, swap the output anesthesia tube to the line leading to the sample holder, and secure the mouse to the bed with the bite bar.
6. Apply eye lube to prevent dehydration during the scan.
7. Insert the rectal temperature probe and secure it with a piece of tape to monitor body temperature. Optimal body temperature is 35–38 °C.
8. Secure the respiratory probe on the chest area to monitor respiration. Optimal respiratory rate is 40–60 breaths/min. (Figure 2d)
9. Secure the mouse to the sample holder with surgical tape, passing it through the middle of the body.

Figure 2. Securing fiducials and mouse for imaging. (A) Fiducial (Vivotrax) aliquots in five small tubes. (B) Fiducials secured on the mouse holder with paper tape. Red-marked areas represent each fiducial. (C) A custom-made mouse bed placed on top of the secured fiducials. (D) A mouse secured on the holder ready for scanning.
D. In vivo magnetic particle imaging scanning protocol
1. Once the mouse is ready, open the Momentum Scanner software. A screenshot of the method is given in Figure 3.
2. Fill in the test name and description.
3. Open the standard mode configuration file in Advanced Mode (Supplementary Data 1).
4. Enable both “x” and “z” transmit axes.
5. Input the field of view for “x” and “z” to 6.0 and 12.0, respectively.
6. Input the gradient field to 5.7.
7. Input the RF amplitude to 5.0 for both “x” and “z.”
8. Input the number of angles to 21.
9. Perform the scan.

Figure 3. Screenshot of the scanning parameters. Scanning parameters selected for 3D MPI scan are shown here from the Momentum software screen, as described in the text.
E. In vivo micro-computed tomography
Note: Since the MPI provides only the nanoparticle signal, microCT imaging is needed to co-register with the MPI image to identify the anatomical positions of tumors and metastases.
1. Ensure microCT is calibrated and background measurements are current.
2. If the MPI system is not integrated with microCT (as shown here), after MPI scanning, transfer the holder with the mouse to the IVIS Spectrum CT imager with minimal disturbance.
3. Remove the rectal probe and respiration monitor tube gently without shifting the location of the mouse or fiducials.
4. Carefully remove the bite bar.
5. Turn off the circulating heater water bath and disconnect all water lines.
6. Carefully remove the water line fittings from the sample holder and drain the water into a waste container.
7. Remove the vacuum line fitting while keeping the anesthesia hose connected.
8. Remove the sample holder and place it in the microCT adapter (Figure 4). Wrap the tubes around the holder tightly, disconnect the anesthesia tubing, and place the whole setup in the IVIS Spectrum CT imager for microCT data acquisition.

Figure 4. Mouse ready for microCT imaging
9. Perform a CT scan using the default single mouse scan in the IVIS Spectrum CT System (Supplementary Data 2).
F. Ex vivo magnetic particle imaging scanning protocol
1. After finishing the in vivo MPI and microCT scans, euthanize the mouse and remove relevant organs: lungs, liver, tumor, intestines, lymph nodes, sternum (bone marrow), kidneys, spleen, and a sample of blood. Store the tissues in relevant containment vessels (small tube or clamshell) and place on dry ice.
2. Secure the tissues individually to the sample holder with a piece of surgical tape (Figure 5).

Figure 5. Individual organ scans. The organ of interest, placed in a plastic cassette, is secured on the sample holder of the MPI scanner.
3. Open the Momentum Scanner software.
4. Fill in the test name and description.
5. Open the standard mode configuration file in Advanced Mode.
6. Enable both “x” and “z” transmit axes.
7. Input the field of view for “x” and “z” to 6.0 and 6.0.
8. Input the gradient field to 5.7.
9. Input the RF amplitude to 10.0 for both “x” and “z.”
10. Input the number of angles to 1.
11. Perform the scan.
G. MPI signal calibration scan
1. Prepare a concentration gradient of 5–6 nanoparticle samples, where the maximum concentration is the injection concentration.
2. Secure a single sample to the holder and perform a scan using either the ex vivo or in vivo scan parameters. Repeat for the remaining samples.
3. Measure each sample at both the ex vivo and in vivo scan parameters.
Data analysis
A. In vivo reconstruction
MPI data were reconstructed using a physics-based model of the momentum field-free-line (FFL) imaging system. The reconstruction incorporates the scanner geometry, applied magnetic field gradients, received sensitivity, and the nonlinear magnetic nanoparticle response to construct a forward model relating the nanoparticle distribution to the measured MPI signal [8]. The 3D FFL reconstruction combines compressed raw data from multiple projection angles within a single 3D model-based reconstruction framework [9].
The reconstruction model is solved via a Tikhonov-regularized iterative reconstruction in MATLAB with GPU acceleration. The reconstruction code is provided at https://github.com/toby-sanders/MPI-FFL-reconstruction. The reconstruction parameters, including voxel size, reconstruction field of view, regularization strength, and number of iterations, are specified near the beginning of the supplied “demo_3D_recon.m” script. Approximately 150–250 iterations are recommended for high-quality reconstruction, although fewer iterations may be used for testing. After reconstruction, maximum-intensity projections (MIP) or reconstructed image volumes (Figure 6) may be exported for subsequent registration with microCT, segmentation, and quantitative analysis as described below. For further mathematical details of the MPI forward model and reconstruction algorithm, refer to [8,9].

Figure 6. MPI-enabled detection of tumor and metastases in mice. Bioluminescent imaging (BLI) shows the primary tumor on the right fourth mammary gland area, with metastases to the left lower leg region and in the lung area of 4T1 tumor-bearing mice. Maximum-intensity projection of the MPI scan of the same mouse shows nanoparticle accumulation in the primary tumor area (dotted yellow circle) and metastatic nodules in the lung and left leg area (orange arrows), as seen in BLI. Fiducials are marked with white arrows.
B. In vivo quantitation
1. Open the image analysis software and load the MPI and microCT DICOM files.
2. Perform a fiducial registration. Identify the fiducial markers on each respective image and perform a rigid transformation of the MPI scan to overlay it onto the microCT scan.
3. Using manual segmentation tools (e.g., Paint or Draw in MagImage) in the software, segment a ~35 mm3 spherical portion of the tumor and thigh muscle (Figure 7). Try to select the region with the highest signal in these areas.
4. Record the MPI signal statistics from within these segmentations, such as the sum of the scalar values, maximum voxel value, and average voxel value.

Figure 7. Manually segmented area for magnetic particle imaging (MPI) signal quantitation. Red circle: tumor; violet circle: muscle.
C. Ex vivo quantitation
1. Open the image analysis software and load the single DICOM file for each tissue.
2. Segment the entire scan area, which should be a 241 × 241-pixel area (58,081 total pixels).
3. Record the MPI signal statistics from within this segmentation.
D. MPI calibration curve creation
1. Open the image analysis software and load the DICOM files for the calibration scans.
2. Segment the entire scan area and record the MPI signal statistics.
3. Plot two different graphs: Maximum MPI signal vs. nanoparticle concentration, and sum of scalar values (total MPI signal) vs. nanoparticle mass. Perform a least-squares fit and record the resulting equation and R2 value.
Validation of protocol
This protocol was followed to detect primary and/or metastatic tumors in multiple tumor models as reported by Korangath et al. [5,6]. All raw data can be accessed through the Johns Hopkins Research Data Repository: https://doi.org/10.7281/T1QZ3N5L and https://doi.org/10.7281/T1DADMSK.
General notes and troubleshooting
General notes
The concentration (dose) of NP and time after injection for the MPI scan described here are based on our research article [5]. The concentration of NP can be reduced, and MPI can be conducted after a shorter interval post-injection. We waited for 72 h for complete clearance of NPs from the bloodstream so that NPs accumulated in small micrometastases could be detected. Studies are currently underway to evaluate an optimal dose and time point.
Troubleshooting
Problem 1: MPI temperature error.
Possible cause: High amounts of nanoparticles can cause the MPI detector to become saturated. During prolonged imaging sessions, such as during the 3D in vivo scans (which can take 30–45 min per scan), this can cause the scanner to overheat.
Solution: Reduce the injected nanoparticle dose or adjust the scan parameters, such as decreasing the RF amplitude or total scan time.
Problem 2: MATLAB reconstruction code not finishing.
Possible cause: Not enough GPU VRAM.
Solution: Try the reconstruction again using a computer containing a GPU with at least 16 GB of VRAM.
Problem 3: Unable to find fiducials during co-registration.
Possible cause: The signal from the fiducials is substantially lower than the nearby sample and unable to be discerned during co-registration.
Solution: Increase the fiducial concentration or adjust the placement of the fiducials away from high-intensity areas in the upper abdomen (e.g., liver and spleen).
Supplementary information
The following supporting information can be downloaded here:
1. Supplementary Data 1.
2. Supplementary Data 2.
Acknowledgments
The authors thank Dr. Olivia C Sehl and Dr. Abdul Rahman Mohtasebzadeh (Magnetic Insight Inc., Alameda, California, USA) for helpful discussions and guidance on MPI; and Dr. Adnan Bibic for instrument support at the Preclinical Imaging Facility, F. M. Kirby Research Center, Kennedy Krieger Institute, Baltimore. This study was funded by Jayne Koskinas Ted Giovanis (JKTG) Foundation for Health and Policy (P.K., R.I.); National Institutes of Health grant R01 CA257557 (R.I., P.W.G.); National Institutes of Health grant R44 CA285064 (P.W.G., R.I.); National Institutes of Health grant R01 CA247290 (R.I.); National Institutes of Health grant R50 CA305053 (P.K.). Magnetic Particle Imaging scanner core, Animal, biostatistics, and other resource facilities were supported in part by National Institutes of Health grants S10 OD026740 and P30 CA006973. The content is solely the responsibility of the authors and does not necessarily represent the official view of the National Institutes of Health, JKTG Foundation or Johns Hopkins University. Funders had no role in the writing of the manuscript or in the decision to publish. This manuscript was funded in part by the National Institutes of Health (NIH). It is subject to the NIH Public Access Policy. Through acceptance of this federal funding, NIH has been given the right to make this manuscript publicly available in PubMed Central upon the Official Date of Publication, as defined by NIH. The Graphical overview was created with BioRender.com under a license (https://BioRender.com/ij5ynbu).
Author contributions
Protocol Development: P.K., H.C., T.S., P.W.G., R.I. Protocol Optimization and Validation: P.K., H.C., T.S. Writing—Original Draft: P.K., H.C., T.S. Writing—Review & Editing: P.K., H.C., T.S., P.W.G., R.I. Funding acquisition: P.K., P.W.G., R.I. Supervision: P.K., P.W.G., R.I.
Competing interests
T.S. and P.W.G. were employees at Magnetic Insight, Inc., C.A. R.I. is an inventor listed on several nanoparticle patents. All patents are assigned to Johns Hopkins University or Aduro Biosciences, Inc. All other authors declare they have no conflicts of interest.
Ethical considerations
All animal studies were performed according to the NIH guidelines for the care and use of laboratory animals [Guide for the Care and Use of Laboratory Animals, 8th edition, National Research Council (US) Committee for the Update of the Guide for the Care and Use of Laboratory Animals. Washington (DC): National Academies Press (US); 2011]. The Institutional Animal Care and Use Committee at Johns Hopkins University approved all animal studies under protocol number M024M16.
References
Article Information
Publication history
Received: Sep 21, 2026
Accepted: Nov 15, 2026
Available online: Nov 27, 2026
Published: Jan 5, 2027
Copyright
© 2027 The Author(s); This is an open access article under the CC BY-NC license (https://creativecommons.org/licenses/by-nc/4.0/).
How to cite
Korangath, P., Carlton, H., Sanders, T., Goodwill, P. W. and Ivkov, R. (2027). Detecting Metastases by Magnetic Particle Imaging Using Pegylated Synomag Nanoparticles in Murine Breast Cancer Models. Bio-protocol 17(1): e5911. DOI: 10.21769/BioProtoc.5911.