(*Contributed equally to this work, §Deceased) Published: Vol 16, Iss 24, Dec 20, 2026 DOI: 10.21769/BioProtoc.5903 Views: 20
Reviewed by: David PaulYuanmin ZhengAnonymous reviewer(s)
Abstract
Plaque assays remain a central method for quantifying infectious viral particles and are frequently used to support potency assessment of live viral vaccines. For recombinant viral vaccine candidates, assay performance must be demonstrated with sufficient specificity, accuracy, precision, linearity, range, and robustness to satisfy regulatory expectations. This protocol describes a Vero E6 cell-based plaque assay for determining the infectious titer of rVSV-SARS-CoV-2-S vaccine preparations, according to the European Medicines Agency (EMA) Good Manufacturing Practice (GMP) regulations. Samples are serially diluted, inoculated onto confluent Vero E6 monolayers, overlaid with tragacanth-containing medium to restrict viral spread, incubated for plaque formation, stained with crystal violet, and counted manually according to defined acceptance criteria. The assay is designed around three endpoint dilutions targeting approximately 15, 30, and 60 plaque-forming units (PFU) per well, with valid countable wells in the 10–80 PFU range. The protocol provides a practical framework for routine titer determination and for analytical validation of plaque assays intended for live virus vaccine preparations.
Key features
• Quantification of infectious rVSV-SARS-CoV-2-S particles in vaccine drug substance (DS) and drug product (DP) samples using Vero E6 cells.
• Use of three endpoint dilutions targeting 15, 30, and 60 PFU/well within a validated 10–80 PFU/well countable range.
• GMP-compatible acceptance criteria for sample, QC, and negative-control validity.
• A validation framework for specificity, linearity, range, accuracy, precision, LLOQ/LOD, and robustness.
Keywords: PFUGraphical overview
Validated plaque assay workflow for rVSV-SARS-CoV-2-S titer quantification. Vero E6 cells are seeded (-24 h), viral endpoint dilutions are prepared (-30 min), cells are inoculated (time 0), the tragacanth overlay is added (at 1 h post-inoculation), cells are stained, and plaques are counted (72 h). The titer in PFU/mL is then calculated. QC and negative controls are included in each assay to verify run validity. The graphical abstract was generated iteratively with the assistance of an AI language model (Gemini, Google). The authors directed the visual generation process, reviewed all iterations, and take full responsibility for the final scientific accuracy of the image.
Background
The rVSV-SARS-CoV-2-S vaccine candidate, BriLife®, was developed by the Israel Institute for Biological Research (IIBR) during the COVID-19 pandemic, which caused extensive global morbidity and mortality [1]. BriLife® is a recombinant vesicular stomatitis virus (VSV)-based vaccine in which the VSV glycoprotein was replaced by the SARS-CoV-2 spike protein; its development, production, purification, toxicology, and protective efficacy have been described [2–6]. The platform was based on the prior successful use of an rVSV-vectored vaccine for Ebola virus disease [7,8] together with proven use of VSV as a vaccine-vector backbone for outbreak pathogens [9].
Traditional plaque assays quantify infectious virus based on the formation of a localized zone of cell destruction initiated by a single infectious viral particle in a confluent layer of cells; each infectious viral particle produces a plaque—a circular zone of lysed or dead cells resulting from viral infection, surrounded by an intact monolayer of uninfected cells that becomes large enough to be visible to the naked eye. The assay outcome is given as plaque-forming units per milliliter (PFU/mL) [10]. For live-virus vaccines, infectious titer assays are used for in-process control, assessment of batch-to-batch consistency, and characterization of drug substance (DS, bulk-purified viral harvest) and drug product (DP, final formulated vaccine vial). They are also used for batch release and potency-related decision-making [11]. In the IIBR BriLife® production program, validated plaque assay titers supported dose determination, and a related plaque reduction neutralization test was used to measure neutralization of SARS-CoV-2 variants in vaccinee sera [12,13].
Assay validation was therefore designed to show that the analytical procedure is suitable for its intended use. The workflow follows the European Pharmacopoeia framework for PFU determination of smallpox vaccine and the ICH Q2(R1) principles for validation of analytical procedures [14,15]. In our original validation study, the assay demonstrated specificity, linearity, range, accuracy, repeatability, intermediate precision, lower limit of quantitation, limit of detection, and robustness across Vero E6 passage numbers [12].
Plaque assays have been used in many viral systems, including mumps virus, alphaviruses, measles virus, rubella virus, SARS-CoV-2, and vaccinia-related neutralization assays [16–21]. Moreover, two previous quantitative plaque-assay validation studies have been published, based on the United States Pharmacopeia (USP) [22,23] and the European Pharmacopoeia and the ICH principles for validation [12,14,15]. However, none of the studies described the plaque assay protocol in detail.
This detailed protocol can serve as a practical framework for conducting plaque assays for other plaque-forming live-virus vaccine candidates, based on our previous study [12]. The protocol enables others to develop and establish plaque assays while highlighting the labor-intensive nature of the assay and its dependence on careful monolayer preparation, appropriate overlay conditions, and well-trained analysts.
Materials and reagents
Biosafety: The containment level and approvals must be determined by the local biosafety committee for the specific viral construct and sample type. This protocol involves infectious recombinant viral material. Perform all work only under institutionally approved biosafety conditions, in a certified biosafety cabinet, with trained personnel, validated decontamination procedures, and approved waste handling.
Biological materials
1. rVSV-SARS-CoV-2-S vaccine samples (IIBR recombinant VSV vector expressing SARS-CoV-2 spike [13]). Storage/notes: Drug substance (DS), drug product (DP), or QC sample; dispensed as single-use aliquots (1.0 mL) and stored at -70 °C in the validated workflow. Samples that were thawed at room temperature must not undergo repeated freeze/thaw cycles.
2. Vero E6 cells (ATCC, origin/catalog: C1008; lot number: 70034202). Storage/notes: Master and Working Cell Banks (WCBs) stored in liquid nitrogen vapor-phase; thaw WCB vials rapidly in a 37°C water bath (1–2 min) and subculture at least twice before assay use. Use approved passage range, validated here at passages 5–19.
3. DP_low sample (IIBR drug product preparation). Storage/notes: Predetermined titer 1.17 × 107 PFU/mL*.
4. DP_ high sample (IIBR drug product preparation). Storage/notes: Predetermined titer 1.60 × 108 PFU/mL*
5. QC sample (IIBR drug product preparation). Storage/notes: Predetermined titer 6.50 × 107 PFU/mL*
*Established across multiple independent qualification runs using this plaque assay [12]; stored at -70 °C in single-use 1.0 mL aliquots.
Reagents
1. Dulbecco’s modified Eagle’s medium (DMEM) (Sartorius, catalog number: 01-055-1A); used for Vero E6 cell culture medium (store at 2–8 °C; use within manufacturer expiry and stability window after opening)
2. Eagle’s minimal essential medium (MEM) (Sartorius, catalog number: 01-035-1A); used for viral dilutions and overlay medium base (store at 2–8 °C)
3. Fetal bovine serum (FBS), heat-inactivated (Sartorius, catalog number: 04-121-1); used for media supplement: 10% in culture medium; 2% in overlay medium (store at -20 °C and at 2–8 °C after thawing; qualify new lots for Vero E6 growth against an approved reference lot)
4. L-glutamine (Sartorius, catalog number: 03-020-1A/B/C); used for media supplement; final concentration 2 mM
5. Nonessential amino acids (NEAA) (Sartorius, catalog number: 01-340-1B); used for media supplement; final concentration 1%
6. Tragacanth gum (Merck, catalog number: G-1128); used for overlay medium; final concentration 0.38% w/v (fine powder stored dry at room temperature; new lots are qualified after ethanol wash with QC sample prior to routine use)
7. Sodium bicarbonate (Sartorius, catalog number: 03-040-1A/B/C); used for overlay medium; final concentration 0.15%
8. Penicillin-streptomycin (Pen-strep) solution (Sartorius, catalog number: 03-031-1A/B/C); used for media supplement; final concentration 0.5% (store stock at -20 °C; store thawed working aliquots at 2–8 °C)
9. Crystal violet dye (Sartorius, catalog number: 01-986-1A); used for staining solution; supplied as a ready-to-use 0.1% w/v aqueous solution (store at room temperature in a light-protected bottle)
10. Tris(hydroxymethyl)aminomethane (Tris) (Merck, catalog number: 1.08386); used for DS/DP formulation buffer (see Recipe 4)
11. Recombinant human serum albumin (rHSA) (Invitria, catalog number: 777HSA097); used for DS/DP formulation buffer (see Recipe 4)
12. NaCl (Merck, catalog number: 1.16224); used for DS/DP formulation buffer (see Recipe 4)
13. Trehalose (Pfanstiehl, catalog number: T-104-4); used for DS/DP formulation buffer (see Recipe 4)
14. Trypsin-EDTA (Sartorius 03-079-1A); used for DS/DP formulation buffer (see Recipe 4)
Solutions
1. Complete Vero E6 culture medium (see Recipes)
2. Virus dilution medium (MEM) (see Recipes)
3. Tragacanth overlay medium (see Recipes)
4. DS/DP formulation buffer (see Recipes)
Recipes
1. Complete Vero E6 culture medium
| Reagent | Final concentration | Quantity or volume (per 100 mL medium) |
|---|---|---|
| DMEM | Base medium | To final volume |
| FBS | 10% | 10 mL |
| L-glutamine | 2 mM | 1 mL of 200 mM stock |
| NEAA | 1% | 1 mL of stock |
| Pen-Strep solution | 0.5% | 0.5 mL |
Store according to the manufacturer and institutional standard operating procedure (SOP). Warm only the required volume before use.
2. Virus dilution medium
| Reagent | Final concentration | Quantity or volume (per 500 mL medium) |
|---|---|---|
| MEM | Base medium | To final volume |
| FBS | 2% | 10 mL |
| L-glutamine | 2 mM | 5 mL of 200 mM stock |
| NEAA | 1% | 5 mL of stock |
| Pen-Strep solution | 0.5% | 2.5 mL |
3. Tragacanth overlay medium
| Reagent | Final concentration | Volume (per 1 L overlay medium) |
|---|---|---|
| MEM X2 | Base medium | 465 mL |
| Tragacanth X2 (0.8% w/v in water) | 0.38% w/v | 470 mL |
| FBS | 2% | 20 mL |
| L-glutamine (200 mM) | 2 mM | 10 mL of 200 mM stock |
| NEAA | 1% | 10 mL of stock |
| Sodium bicarbonate (7.5%) | 0.15% w/v | 20 mL |
| Pen-Strep solution | 0.5% | 5 mL |
Critical: Prepare a homogeneous tragacanth-containing overlay to ensure reproducible plaque morphology, as uneven viscosity can alter plaque size and plaque distribution.
Prior to dissolution in water, subject the tragacanth powder to the following washing procedure: Wash the tragacanth three times with absolute ethanol at a 1:3 ratio (w/v, tragacanth:ethanol). For each wash, stir the suspension using a magnetic stirrer for 5–7 min to ensure complete mixing of the powder, then stop stirring and allow the tragacanth to settle. Following the third wash, remove the ethanol by filtration, spread the recovered tragacanth on aluminum foil, and dry it at 37 °C for several hours until completely dry. After drying, collect the powder into a sterile, tightly sealed bottle and store at room temperature in a dry environment for up to 1 year. To prepare the 2× (0.8% w/v) stock, dissolve the washed powder in sterile water and sterilize by autoclaving before storing at 2–8 °C (or room temperature) until use, for up to 1 year.
4. DS/DP formulation buffer
| Reagent | Quantity or volume |
|---|---|
| Tris | 2.4228 g per 1 L |
| rHSA | 2.5 g per 1 L |
| NaCl | 9 g per 1 L |
| Trehalose | 40 g per 1 L |
Note: To prepare the formulation buffer, dissolve Tris (20 mM), NaCl (154 mM), and trehalose (4% w/v) sequentially in water for injection (WFI) with magnetic stirring, adjust/verify physiological pH, gently add rHSA (0.25% w/v) without foaming, bring to 1 L final volume, sterilize by 0.22 μm filtration, and store at 2–8 °C. In this protocol, the formulation buffer serves as the virus-negative matrix control (blank) and for virus-spiking studies during specificity validation (see Section H), whereas routine serial dilutions for plaque titration are performed in virus dilution medium (Recipe 2).
The formulation buffer composition should match the approved manufacturing record and should not be changed for routine sample dilution unless revalidated. Users adapting this protocol to another virus or vaccine candidate should use their own product-specific formulation buffer for specificity/matrix validation while using the supplemented MEM virus dilution medium (Recipe 2) for routine serial dilutions.
Laboratory supplies
1. 6-well tissue-culture plates
2. Sterile conical tubes and microcentrifuge tubes compatible with live-virus work
3. Sterile serological pipettes and filtered pipette tips
4. Single-use frozen sample ampoules/vials for DP, DS, or QC samples
5. Paper or electronic batch record according to GMP regulations
6. Locked and secured Microsoft Excel spreadsheet, according to GMP regulations
Equipment
| Equipment | Manufacturer/model | Use |
|---|---|---|
| Class II biosafety cabinet | Any | Required for infectious material handling and for sterility |
| CO2 incubator for cell culture | Any | 37 ± 2 °C, 5% CO2, humidified atmosphere. Position on a stable, vibration-free surface (away from centrifuges/vortexers) and dedicate an undisturbed incubator during the 72 h incubation |
| Inverted microscope | Eclipse Ts2R, Nikon Instruments Inc. | Equipped with 4× and 10× phase-contrast objectives for cell confluency and monolayer inspection |
| Hand-held colony counter pen | SP Bel-Art hand-held colony counter, catalog number: F37862-0000 | Manual plaque counting (marks each plaque on the plate bottom upon contact; an illuminated light box or magnifying glass may be used to aid visualization). |
| -70 °C freezer | Any | Storage of aliquoted viral samples |
| Liquid nitrogen vapor-phase storage system | Any | Storage of master cell bank (MCB)/working cell bank (WCB) vials |
| Cell-counting instrument | Any | Required for cell seeding |
| Wave rocker, placed inside the CO2 incubator | Any | Required for continuous gentle back-and-forth rocking of plates during the 1 h viral adsorption step |
Software and datasets
| Software/dataset | Version/source | Use |
|---|---|---|
| Microsoft Excel | Locked and secured Microsoft Excel spreadsheet. | Routine mean, SD, and titer calculations |
| GraphPad Prism | GraphPad Prism 5 | Regression lines, ANOVA, correlation analysis, and figure generation during validation |
| Batch records/SOPs | Institution-specific controlled documents | Required for GMP execution and traceability |
Procedure
A. Prepare Vero E6 cell monolayers
1. Maintain Vero E6 cells from the approved Working Cell Bank (thawed rapidly at 37 °C and subcultured at least twice) by washing the monolayer with 1× PBS, detaching cells with trypsin-EDTA for 5 min at 37 °C, neutralizing with complete Vero E6 culture medium (Recipe 1), and splitting at a 1:3–1:6 ratio no more than two days before assay seeding (see Supplementary Protocol 1). Use only cell passages within the validated range (passages 5–19 [12]) and document the passage number for each assay in the batch record.
Caution: Using cells outside the validated passage range may alter cell growth rate, monolayer integrity, or viral susceptibility, leading to assay failure or inaccurate titers.
2. One day before infection, prepare a Vero E6 cell suspension in complete Vero E6 culture medium at 700,000 ± 50,000 cells per 3 mL (~2.33 × 105 cells/mL; this cell number per well was optimized during method development to reliably produce a 90%–100% confluent monolayer within 22–30 h).
3. Slowly seed 3 mL of the cell suspension into each well of a sterile 6-well tissue-culture plate (700,000 ± 50,000 cells/well), gently mixing the cell suspension tube before each pipette aspiration. Prepare enough wells for all test sample dilutions, QC sample wells, and negative-control wells. Since a confluent monolayer is critical, if necessary, gently agitate the plate after 30 min to ensure that the cells are evenly distributed throughout the well.
4. Incubate the plates for 22–30 h at 37 ± 2 °C and 5% CO2. Since an even, confluent monolayer is critical, gently slide each plate manually side-to-side and back-and-forth on a flat surface (do not swirl in circles) immediately after seeding and again 30–60 min after placing the plates in the incubator.
5. Before virus inoculation, inspect each well under an inverted microscope (4× or 10× objective). While plates may be transferred from the cell-culture preparation room once they reach ≥80% confluency (Supplementary Protocol 1), proceed with viral inoculation only in wells that have reached an intact, 90%–100% confluent monolayer with typical morphology and without cell clumping, microbial contamination (turbidity or pH change), or gross mechanical damage (see Figure 1).
Critical: Monolayer quality is a major source of variability. Damaged, non-confluent, or overgrown wells should not be used for plaque counting. Document any excluded wells and the reason for exclusion in the batch record.

Figure 1. Four cell densities and their suitability for the assay. (A) A subconfluent cell layer (~40%–50% confluent) that has not reached the confluence required for the assay. (B) A confluent monolayer (90%–100% confluent, suitable for the assay), with a typical triangular cellular morphology and no cell overlap. (C) Over-confluent cells (~110% confluent, acceptable but suboptimal for assay initiation), with a hexagonal morphology and cells beginning to overlap in several areas. (D) Highly over-confluent cells (>120% confluent, unsuitable for the assay). The cells have lost their natural morphology and display multi-layered growth.
B. Prepare viral sample dilutions
1. Remove single-use frozen vials of the test sample and QC sample from the -70 °C freezer and thaw at room temperature until completely liquid (do not refreeze). Mix the thawed stock gently by inversion or gentle swirling (5–10 times) to avoid foaming or excessive shear (see Supplementary Protocol 2).
2. Prepare tenfold serial dilutions of the rVSV-SARS-CoV-2-S sample in virus dilution medium (supplemented MEM; see Recipe 2), such as adding 0.1 mL of virus suspension to 0.9 mL of dilution medium (or 0.2 mL into 1.8 mL when larger volumes are needed; see Tables 1 and 2 in Supplementary Protocol 2). Mix tubes briefly by vortexing between serial dilution steps.
3. Based on the predetermined or expected titer, first complete the tenfold serial dilution series to reach the working concentration range and then prepare the three final twofold endpoint dilutions of 300, 150, and 75 PFU/mL from the appropriate tenfold dilution tube (see Table 2 in Supplementary Protocol 2 for the exact dilution scheme). When 0.2 mL is inoculated per well, these endpoint dilutions correspond to approximately 15, 30, and 60 PFU/well, respectively.
4. Prepare each endpoint dilution in a sufficient volume for six replicate wells per dilution, plus the excess volume required by the local SOP for pipetting dead volume. For example, for six replicate wells of 0.2 mL each (1.2 mL total), prepare at least 1.4–2.0 mL of each final dilution (see Table 2 in Supplementary Protocol 2).
5. Prepare negative-control wells using dilution medium without virus.
Note: For samples with an undetermined titer, an initial pilot assay spanning a broad tenfold dilution series should be performed first to identify the appropriate working range, so that at least one endpoint dilution falls within the valid countable range of 10–80 PFU/well.
C. Inoculate Vero E6 monolayers
1. Tilt each 6-well plate slightly and gently aspirate the culture medium from the side wall/lowest corner of the wells designated for infection without touching or disrupting the monolayer, leaving a minimal residual liquid film (~0.1 mL) so that cells do not dry out prior to inoculation.
2. Gently add 0.2 mL of each endpoint dilution down the side wall of each well of a 6-well plate (avoid pipetting directly onto the cell sheet to prevent mechanical detachment) and gently tilt the plate so the inoculum covers the monolayer. Inoculate six replicate wells for each dilution.
3. Add 0.2 mL of virus-free dilution medium to each negative-control well.
4. Incubate the plates for 1 h at 37 ± 2 °C and 5% CO2 to allow viral adsorption.
Critical 1: Do not allow cell monolayers to dry during aspiration and inoculation. Process only a small batch of plates (e.g., 2–3 plates) at a time and keep plate lids closed whenever possible so wells are exposed to cabinet airflow for no more than a few minutes. Drying can cause cell damage that mimics plaques or increases background artifacts.
Critical 2: To ensure homogeneous virus distribution and keep the monolayer moist throughout the entire 1-h adsorption period, place the plates on a continuous slow wave rocker (gentle back-and-forth rocking at ~10–15 cycles/min, ~5–10° tilt angle) inside the humidified CO2 incubator. Do not use an orbital shaker.
D. Add tragacanth overlay and incubate for plaque development
1. After the 1-h adsorption step, do not remove the 0.2 mL viral inoculum. Gently dispense 3 mL of prewarmed (room temperature to 37 °C) tragacanth overlay medium (Recipe 3) down the side wall of each well directly over the inoculum (resulting in a final total volume of ~3.2 mL per well).
2. Incubate plates for 72 h at 37 ± 2 °C and 5% CO2. During this period, the tragacanth overlay restricts viral spread to neighboring cells, allowing discrete plaques to form.
Critical: The incubator must not be disturbed during this period; even opening the door can cause vibrations that disrupt plaque formation. It is recommended to use a dedicated incubator for plaque assays and post "Do not open/Vibration-sensitive assay in progress" signage during the 72-h incubation.
E. Stain and count plaques
1. At the end of the 72-h incubation, transfer the plates to a Class II biosafety cabinet and slowly aspirate the viscous tragacanth overlay from the tilted corner of each well into a disinfectant-containing vacuum trap without touching the cell monolayer.
2. Add 1 mL of 0.1% crystal violet staining solution to each well and incubate for 4–6 min at room temperature in the biosafety cabinet (this window ensures strong contrast between stained viable cells and unstained plaques while avoiding nonspecific staining of cellular debris within plaques).
3. Aspirate the staining solution into the disinfectant trap, gently rinse each well 1–2 times by adding ~2–3 mL of tap water down the side wall of the well inside the biosafety cabinet, aspirate the rinse water into the disinfectant trap, and allow the plates to air dry.
4. Invert the plates and count visible plaques manually using a hand-held colony counter pen (marking each counted plaque on the exterior bottom of the well; dried stained plates may be counted immediately or stored dry at room temperature for a few days). Any distinct macroscopic unstained circular clearing is counted as one plaque; when two adjacent plaques partially merge, count them as separate plaques if their individual circular centers/contours remain clearly distinguishable (see Figure 2A). Record plaque counts for each well directly in the controlled batch record and in the locked calculation spreadsheet.
5. Exclude wells containing fewer than 10 PFU/well (marked TL, too low) or more than 80 PFU/well (marked TH, too high) from the titer calculation (see Figure 2). In addition, for intraplate consistency, exclude individual outlier wells whose plaque count falls outside 70%–130% (±30%) of the initial plate mean (up to two outlier wells may be omitted per 6-well plate, after which the plate mean is recalculated from the remaining ≥4 wells; see Recipe 1). If the plate %CV still exceeds 30% after outlier removal, exclude that dilution plate from the calculation (see Supplementary Protocol 2, Section K).

Figure 2. Plaque formation on the cell monolayer following crystal violet staining. (A) A valid assay well (containing 62 PFU/well): distinct, countable plaques of varying sizes are visible on the cell monolayer (green arrows indicate representative small and large individual plaques). Even when adjacent plaques partially merge, their individual circular centers remain distinguishable and countable (green dashed circles). The total plaque count falls within the acceptable quantitation range (10–80 PFU/well). (B) Invalid assay well: Plaques are excessively large and overlapping (red dashed outlines) and cannot be individually resolved, precluding accurate enumeration. (C) Invalid negative control: Detachment of the cell monolayer is visible along the well edge (red dashed outline), which can result from mechanical scraping during pipetting/aspiration, monolayer drying during inoculation, harsh washing, or over-confluent/aged cells. Areas where cells have been removed may lead to confusion during plaque counting on the assay plates.
F. Calculate sample titer
Calculate the titer for each valid dilution as follows:
PFU/mL = mean PFU per well × dilution factor × 5
The factor 5 normalizes the inoculated volume from 0.2 mL per well to 1 mL. Calculate the mean titer from the valid wells and replicate plates according to the rules below.
1. For each sample vial, calculate the titer from valid wells across the endpoint dilutions. In the published validation [12], each dilution was tested in six replicate wells.
2. For DP sample testing, use three independently thawed frozen vials when performing the full validated assay format. Calculate the final sample concentration from the average of the three vials, at two different time points (assays). The final result will be calculated as the average of the results obtained at the two time points.
3. For each vial, calculate the average from at least eight valid replicate results across the tested dilutions (from two different dilutions, with at least four wells per dilution).
4. For QC determinations, test three independent replicates of a single dilution, each in six replicate wells, according to the validation workflow.
5. Report titers in PFU/mL and, where needed for validation, as log10 PFU/mL.
Summary of replicate structure and minimum requirements for a valid result:
a. Per dilution (plate): Each endpoint dilution is tested in 6 replicate wells (one 6-well plate). At least 4 valid wells (within 10–80 PFU/well, within 70%–130% of the plate mean, and with plate %CV ≤ 30%) are required for a valid dilution plate.
b. Per sample vial: Each vial is tested across three twofold endpoint dilutions (targeting 15, 30, and 60 PFU/well; 3 plates = 18 wells total). A valid vial titer requires at least two valid dilutions and at least eight valid replicate wells in total across those dilutions.
c. Per DP sample batch determination: For full GMP DP testing, three independently thawed frozen vials are tested in each assay (9 plates = 54 wells per assay), and the assay is performed at two independent time points (two separate assay runs conducted on different days, at least 24 h apart and both initiated within 1 week of freezing the DP batch). Calculate the final DP concentration as the average of the three vials at each time point, and then as the grand average of the two time points.
d. Per QC sample determination: Test three independent replicates of a single target dilution, each inoculated into six replicate wells (3 plates = 18 wells total; at least four valid wells required per plate).
G. Run validity and acceptance criteria
1. QC or negative control failure: If either the negative cell control or the QC sample fails to meet the acceptance criteria in Table 1, the entire assay run is invalid. All test samples in that run must be rejected and re-assayed in a new independent run, even if individual test samples met their internal precision criteria.
Table 1. Validity and acceptance criteria for the plaque assay
| Parameter | Acceptance criterion |
|---|---|
| Cell control | The uninfected cell monolayer remains intact, and no plaques are visible |
| Countable wells | Only wells with 10–80 PFU/well and within 70%–130% (±30%) of the plate mean are included in calculations (minimum 4 valid wells per plate and plate %CV ≤ 30%) |
| Combined sample concentration | 95% confidence interval for the combined concentration from three vials is not greater than ±0.5 log10 PFU/mL |
| QC sample precision | 95% confidence interval for the estimated QC concentration from three replicates is not greater than ±0.5 log10 PFU/mL |
| QC sample accuracy | QC titer differs by less than 0.5 log10 PFU/mL from its established value |
| Cell passage | Use Vero E6 cells only within the validated passage range |
2. Test sample failure in a valid run: If the negative cell control and QC sample pass, the assay run is valid. If an individual test sample fails to meet its internal acceptance criteria (e.g., fewer than 8 valid countable wells across at least 2 dilutions per vial, plate %CV > 30%, or 95% CI > ± 0.5 log10 PFU/mL), only that specific test sample needs to be repeated.
3. Retesting limits: Under GMP regulations, any invalid assay or out-of-specification (OOS) result must be documented and investigated according to the institutional OOS/invalid-assay SOP, which governs root-cause investigation and restricts the maximum number of permitted repeat assays.
H. Optional validation workflow for new samples or adapted plaque assays
When adapting this protocol to another live viral vaccine or a different cell/virus system, validate the method before routine use. The following workflow mirrors the validated rVSV-SARS-CoV-2-S plaque assay.
1. Specificity: Test virus-negative process and formulation matrices (e.g., DS/DP formulation buffer, Recipe 4), with and without known process-related impurities or excipients, and compare them with virus-spiked positive matrices.
2. Linearity: Test a representative low-titer DP sample in three endpoint dilutions targeting 15, 30, and 60 PFU/well. Assess regression of PFU/well vs. reciprocal dilution.
3. Range: Confirm that precision and accuracy are acceptable across the intended countable range. In [12], the practical range was 10–80 PFU/well, with targeted assay design at 15–60 PFU/well.
4. Accuracy: Compare measured titers with predetermined titers for representative and DP samples.
5. Precision: Assess repeatability within assays and intermediate precision across days, analysts, and independent assays.
6. Limit of detection (LOD) and lower limit of quantification (LLOQ): Define the visual limit of detection and the lower limit of quantitation. In this assay, LOD was one PFU/well and LLOQ was 10 PFU/well.
7. Robustness: Test defined sources of normal variation, such as cell passage number. In [12], passages 5–19 did not meaningfully affect titer.
Data analysis
Use validated spreadsheets or controlled statistical-analysis files according to GMP regulations. Raw plaque counts should be transcribed from the batch record and independently verified according to institutional data-integrity procedures. Data analysis was performed using the statistical metrics described in Table 2.
Table 2. Calculations and statistical metrics used for plaque assay data analysis.
| Metric | Calculation or use |
|---|---|
| Mean PFU/well | Average plaque count from all valid wells at the same dilution or across validated dilution/replicate structure |
| PFU/mL | mean PFU/well × dilution factor × 5 |
| Log10 titer | log10 (PFU/mL) |
| Coefficient of variation (%CV) | (Standard deviation/mean) × 100 |
| Relative error (%RE) | [(mean counted PFU - expected PFU)/expected PFU] × 100 |
| 95% confidence interval | mean log10 titer ± t(0.975, n-1) × SD/√n, or the corresponding validated spreadsheet formula |
| Regression analysis | Least-squares regression of PFU/well vs. reciprocal dilution. In validation, R2 ≥ 0.98 was achieved in all nine assays |
| Correlation analysis | Used for robustness assessment of titer vs. Vero E6 passage number |
For validation, test whether dilution affects precision or accuracy using the predetermined statistical plan. In the original validation, protected Microsoft Excel sheets were used for routine calculations, according to GMP regulations. GraphPad Prism 5 was used for regression, ANOVA, correlation analysis, and figure generation [12,15]. Table 3 shows an example calculation of the DP low titer in one plaque assay.
Table 3. Calculation of the DP_low titer in one plaque assay. Three vials were tested in three dilutions. Each dilution was tested in six replicates in a 6-well plate. The factor of five is the quotient of 1 mL divided by 0.2 mL of inoculated virus volume. The lower and upper limits are 70% and 130% of the mean, respectively. The plate titer is the product of the mean plaque count and the total dilution factor. The vial titer is the mean result of three plates. The assay titer is the mean result of three vials.
| Vial 1 | Vial 2 | Vial 3 | |||||||
| Plate 1 | Plate 2 | Plate 3 | Plate 1 | Plate 2 | Plate 3 | Plate 1 | Plate 2 | Plate 3 | |
| Virus dilution (*104) | 3.75 | 7.50 | 15 | 3.75 | 7.50 | 15 | 3.75 | 7.50 | 15 |
| Factor | 5 | 5 | 5 | 5 | 5 | 5 | 5 | 5 | 5 |
| Total dilution (*105) | 1.875 | 3.75 | 7.50 | 1.875 | 3.75 | 7.50 | 1.875 | 3.75 | 7.50 |
| Well 1 | 55 | 27 | 19 | 52 | 27 | 19 | 51 | 30 | 16 |
| Well 2 | 50 | 19^ | 13 | 55 | 29 | 14 | 58 | 31 | 17 |
| Well 3 | 54 | 35 | 19 | 59 | 31 | 20 | 49 | 31 | 18 |
| Well 4 | 52 | 28 | 19 | 51 | 31 | 16 | 46 | 27 | 18 |
| Well 5 | 51 | 30 | 14 | 51 | 34 | 17 | 49 | 31 | 15 |
| Well 6 | 52 | 29 | 12 | 52 | 31 | 8^^ | 43 | 27 | 14 |
| Mean | 52 | 30 | 16 | 53 | 31 | 17 | 49 | 30 | 16 |
| Standard deviation | 1.9 | 3.1 | 3.3 | 3.1 | 2.3 | 2.4 | 5.1 | 2.0 | 1.6 |
| %CV | 3.6 | 10.5 | 20.9 | 5.9 | 7.7 | 13.9 | 10.3 | 6.7 | 10.0 |
| Lower limit | 36.6 | 20.9 | 11.2 | 37.3 | 21.4 | 12.0 | 34.5 | 20.7 | 11.4 |
| Upper limit | 68.0 | 38.4 | 20.8 | 69.3 | 39.7 | 22.4 | 64.1 | 38.4 | 21.2 |
| Plate titer (PFU/mL) | 9.81×106 | 1.12×107 | 1.20×107 | 1.00×107 | 1.14×107 | 1.29×107 | 9.25×106 | 1.11×107 | 1.23×107 |
| Vial titer (PFU/mL) | 1.10×107 | 1.14×107 | 1.09×107 | ||||||
| Assay titer (PFU/mL) | 1.11×107 | ||||||||
^ The plaque count (19) in well 2 was below the lower limit (70% of the mean calculated using the value 19). After omitting the value 19, a new mean based on the remaining five plaque counts was calculated.
^^ The plaque count (8) in well 6 was below the minimum allowed count (10 plaques per well) and was therefore considered too low and omitted from the mean calculation.
Table 4 shows an example of confidence intervals of the mean plaque results in the 6-well plates.
Table 4. Confidence intervals of the mean plaque results in the 6-well plates. The data are analyzed for the DP_low sample in one assay. Well titers are the result of the plaque number multiplied by the dilution. The confidence intervals (CIs) were calculated using a Student’s t-test. Minimum is the Log10(titer – CI). Maximum is the Log10(titer + CI). Length is the difference between the Maximum or the Minimum and the Log10(titer) for the upper or lower 95% CI, respectively. Alpha is the significance level (1 − 0.95).
| Vial 1 | Vial 2 | Vial 3 | |||||||
| Plate 1 | Plate 2 | Plate 3 | Plate 1 | Plate 2 | Plate 3 | Plate 1 | Plate 2 | Plate 3 | |
| Well 1 titer | 1.0×107 | 1.0×107 | 1.4×107 | 9.8×106 | 1.0×107 | 1.4×107 | 9.6×106 | 1.1×107 | 1.2×107 |
| Well 2 titer | 9.4×106 | ^ | 9.8×106 | 1.0×107 | 1.1×107 | 1.1×107 | 1.1×107 | 1.2×107 | 1.3×107 |
| Well 3 titer | 1.0×107 | 1.3×107 | 1.4×107 | 1.1×107 | 1.2×107 | 1.5×107 | 9.2×106 | 1.2×107 | 1.4×107 |
| Well 4 titer | 9.8×106 | 1.1×107 | 1.4×107 | 9.6×106 | 1.2×107 | 1.2×107 | 8.6×106 | 1.0×107 | 1.4×107 |
| Well 5 titer | 9.6×106 | 1.1×107 | 1.1×107 | 9.6×106 | 1.3×107 | 1.3×107 | 9.2×106 | 1.2×107 | 1.1×107 |
| Well 6 titer | 9.8×106 | 1.1×107 | 9.0×106 | 9.8×106 | 1.2×107 | ^ | 8.1×106 | 1.0×107 | 1.1×107 |
| Mean titer | 9.81×106 | 1.12×107 | 1.20×107 | 1.00×107 | 1.14×107 | 1.29×107 | 9.25×106 | 1.11×107 | 1.23×107 |
| Standard deviation | 3.49×105 | 1.17×106 | 2.27×106 | 5.89×105 | 8.79×105 | 1.79×106 | 9.54×105 | 7.41×105 | 1.22×106 |
| Alpha | 0.05 | 0.05 | 0.05 | 0.05 | 0.05 | 0.05 | 0.05 | 0.05 | 0.05 |
| Size | 6 | 5 | 6 | 6 | 6 | 5 | 6 | 6 | 6 |
| CI | 3.66×105 | 1.45×106 | 2.39×106 | 6.18×105 | 9.23×105 | 2.22×106 | 1.00×106 | 7.77×105 | 1.29×106 |
| Log10(titer) | 6.99 | 7.05 | 7.10 | 7.00 | 7.06 | 7.11 | 6.97 | 7.04 | 7.09 |
| Minimum | 6.98 | 6.99 | 7.01 | 6.97 | 7.02 | 7.03 | 6.92 | 7.01 | 7.04 |
| Length (to Min.) | -0.02 | -0.06 | -0.09 | -0.03 | -0.04 | -0.08 | -0.05 | -0.03 | -0.05 |
| Maximum | 7.01 | 7.10 | 7.18 | 7.03 | 7.09 | 7.18 | 7.01 | 7.07 | 7.13 |
| Length (to Max.) | 0.02 | 0.05 | 0.08 | 0.03 | 0.03 | 0.07 | 0.04 | 0.03 | 0.04 |
^ These results were excluded from the analysis, as described in Table 3.
Validation of protocol
This protocol has been used and validated in the following research article:
• Toister et al. [12]. Development and Validation of a Plaque Assay to Determine the Titer of a Recombinant Live-Attenuated Viral Vaccine for SARS-CoV-2. Vaccines (Figures 1–7 and Tables 1–4).
Validation parameters are summarized in Table 5.
Table 5. Summary of plaque assay validation parameters and published performance.
| Validation parameter | Published performance |
|---|---|
| Specificity | Virus-negative DS and DP gave negative results, while virus-spiked matrices yielded expected positive titers [12]. |
| Linearity | Nine independent assays of DP showed excellent linearity with R2 ≥ 0.98 for PFU/well versus dilution [12]. |
| Range | Validated assay design targeted 15–60 PFU/well, while practical countable wells were accepted between 10 and 80 PFU/well [12]. |
| Accuracy | Mean deviations from predetermined titers were 0.01, 0.02, and 0.09 log10 for DP_low, DP_high, and QC samples, respectively [12]. |
| Repeatability | 95% confidence intervals met the European Pharmacopoeia ±0.5 log10 PFU/mL criterion across QC and DP samples [12]. |
| Intermediate precision | Inter-assay %CV values were 12.0% for DP_low, 6.0% for DP_high, and 18.7% for the QC sample [12]. |
| Robustness | Vero E6 passages 5, 10–11, and 16–19 produced titers within 0.095 log10 PFU/mL of the established DP titer; no correlation with passage number was observed [12]. |
The precision of the plaque assay was resolved as repeatability (Table 6) and intermediate precision (Figure 3D).
Table 6. Repeatability of the DP_low titers within one assay. The confidence interval for the assay mean was calculated using a Student’s t-test. Minimum is the Log10(titer - CI). Maximum is the Log10(titer + CI). Length is the difference between the Maximum or the Minimum and the Log10(titer) for the upper or lower 95% CI, respectively.
| Analysis category | Confidence parameter | Calculated value |
| Confidence interval (CI) calculation | Significance level alpha | 0.05 |
| standard deviation | 1.21 × 107 | |
| size | 9 | |
| CI | 9.33 × 105 | |
| Mean assay titer parameters | Titer | 1.11 × 107 |
| Log10(titer) | 7.05 | |
| Lower 95% CI | Minimum | 7.01 |
| Length = Log10(titer - CI) - Log10(titer) | -0.04 | |
| Upper 95% CI | Maximum | 7.08 |
| Length = Log10(titer + CI) - Log10(titer) | 0.04 |

Figure 3. Summary of plaque assay validation parameters (DP low results). (A) Linearity: Regression line for mean PFU/well results vs. dilutions of the DP_low sample. (B) Intra-dilution precision (%CV) of the DP_low sample tested across three twofold dilutions (targeting 15, 30, and 60 PFU/well); bars and error bars represent the mean %CV ± standard deviation (SD) across nine independent validation assays. (C) Intra-dilution accuracy (%RE) across three twofold dilutions of the DP_low sample; bars and error bars represent the mean %RE ± SD across nine independent validation assays. (D) Intermediate precision (inter-assay variation) of DP_low titers across nine independent assays; open circles represent individual assay mean titers, and the horizontal solid blue line and error bars indicate the overall mean titer ± SD. (E) Accuracy and repeatability depicted by the titer distribution of the DP_low sample across each of the nine validation assays; open circles represent individual vial/replicate titers, the horizontal solid blue line represents the predetermined target titer, and the upper and lower red dashed lines represent the European Pharmacopoeia acceptance limits (±0.5 log10 PFU/mL). (F) Robustness of the plaque assay: Mean titers ± SD of the DP_low samples plotted according to the passage number of the Vero E6 cells (passages 5–19).
General notes and troubleshooting
General notes
1. The assay is applicable only to viruses that form visible, countable plaques in the selected cell line and overlay medium.
2. Typical working schedule and assay throughput: Each plaque assay spans a 5-day workflow:
a. Day 1: Harvest Vero E6 cells and seed 6-well plates (22–30 h incubation).
b. Day 2: Thaw samples, prepare serial and endpoint dilutions, inoculate monolayers (1-h adsorption on a wave rocker), and apply tragacanth overlay.
c. Days 2–5: Undisturbed 72-h incubation.
d. Day 5: Aspirate overlay, stain with crystal violet, count plaques manually, and enter data into the validated spreadsheet.
A single qualified analyst can realistically process 1–2 full DP test batches per assay run (12–24 plates total, comprising 9 plates per DP batch, 3 QC plates, and negative control wells).
3. Analyst training is essential because manual plaque counting can introduce subjectivity. To minimize counting subjectivity and handling variability, analysts undergo a structured GMP qualification program prior to routine testing. Training includes supervised cell seeding, serial dilutions, and parallel plaque counting of stained plates alongside a certified analyst. For formal qualification, the analyst must independently perform at least three qualification assays using the QC or reference DP sample, achieving mean titers within ±0.5 log10 PFU/mL of the assigned value and an inter-analyst %CV ≤ 20%. Periodic requalification is maintained through routine QC sample tracking on control charts.
4. The countable range is determined by both biological and visual factors. Counts below 10 PFU/well have high relative counting error, while counts above 80 PFU/well may lead to plaque convergence.
5. Manual plaque counting using a hand-held colony counter pen was validated in this protocol because trained human analysts reliably resolve plaques of varying sizes and partially touching plaques under tragacanth overlays. Automated or semi-automated imaging colony counters may be used to increase throughput provided they undergo formal comparability and bridging validation demonstrating equivalent accuracy, precision, and linearity across the 10–80 PFU/well countable range.
6. Do not change the cell line, passage range, overlay composition, incubation time, staining procedure, or counting criteria without documented comparability or revalidation.
7. All live-virus work must follow locally approved biosafety procedures, including decontamination of liquid waste, plates, and disposable materials.
Troubleshooting (Table 7)
Table 7. Troubleshooting guide for the plaque assay.
| Problem | Possible cause(s) | Solution(s) |
|---|---|---|
| No plaques in positive sample | Incorrect dilution, inactive/thawed sample, failed adsorption, non-permissive or unhealthy cells | Verify sample identity and storage history, repeat with lower dilutions and check incubator conditions. In addition, confirm Vero E6 viability (≥90%) via Trypan Blue exclusion prior to seeding and ensure the reference QC virus sample is included in every run as a positive control. |
| Plaques in negative control | Cross-contamination during dilution or inoculation, contaminated reagents, labeling error | Discard the run, review pipetting workflow and plate map, prepare fresh negative-control medium, and repeat under strict unidirectional handling. Review the plate map for systematic spatial patterns (e.g., splashing or cross-contamination from adjacent high-concentration wells). |
| Too many plaques to count | Endpoint dilution too concentrated | Exclude wells >80 PFU/well and repeat using higher dilution if no valid wells remain. |
| Too few plaques to count | Endpoint dilution too dilute or unexpectedly low sample titer | Exclude wells <10 PFU/well and repeat using lower dilution if no valid wells remain. |
| Diffuse or merged plaques | Overlay not homogeneous, overlay too thin, incubation too long, high plaque density, the incubator was subject to slight vibrations. | Prepare fresh homogeneous tragacanth overlay, verify overlay volume, and ensure endpoint dilutions target 15–60 PFU/well. If incubator vibration is suspected, place a dish of water on the incubator shelf and inspect for surface ripples before starting the next assay. |
| Damaged monolayer after staining | Mechanical disruption during aspiration, drying before overlay, harsh washing | Aspirate gently, avoid drying, and verify staining/washing steps against the approved SOP. |
| High intra-assay variability | Uneven cell seeding, pipetting error, inconsistent adsorption, subjective counting | Mix cell suspension frequently during seeding, use calibrated pipettes, standardize timing, and use trained/certified analysts. |
Supplementary information
The following supporting information can be downloaded here:
Supplementary Protocol 1: Preparation of Vero E6 cell plates for the plaque assay
Supplementary Protocol 2: Plaque assay procedure
Acknowledgments
This protocol draft was adapted from the original publication by Toister et al. [12]. This study was funded by the Israel Institute for Biological Research, grant number SB-127.
Author contributions
Conceptualization, E.T., Y.O., and E.D. (Eran Diamant); Methodology, E.T., L.C., E.L., E.E., Y.O., and E.D. (Eran Diamant); Assay development and validation, L.C., E.L., A.M., E.D. (Eyal Dor), L.L., M.G., N.N., S.C., and S.S.; Formal analysis, E.T., L.C., E.L., Y.A., R.Z., Y.O., and E.D. (Eran Diamant); Writing—original draft preparation, E.T., L.C., E.L., Y.O., and E.D. (Eran Diamant); Writing—review and editing, E.T., Y.A., E.E., R.Z., Y.O., and E.D. (Eran Diamant); Supervision, Y.O. and E.D. (Eran Diamant).
Competing interests
The authors declare no conflicts of interest.
Ethical considerations
Not applicable. The described protocol is an in vitro analytical assay using established Vero E6 cells and viral vaccine preparations. Users must obtain all applicable institutional biosafety approvals before performing the assay with infectious recombinant viral material.
References
Article Information
Publication history
Received: Sep 22, 2026
Accepted: Nov 8, 2026
Available online: Nov 26, 2026
Published: Dec 20, 2026
Copyright
© 2026 The Author(s); This is an open access article under the CC BY-NC license (https://creativecommons.org/licenses/by-nc/4.0/).
How to cite
Toister, E., Cherry, L., Lupu, E., Monash, A., Dor, E., Levin, L., Girshengorn, M., Natan, N., Chapman, S., Shmaya, S., Epstein, E., Adar, Y., Zichel, R., Ophir, Y. and Diamant, E. (2026). A Plaque Assay to Determine the Titer of a Recombinant Live-Attenuated VSV Vaccine for SARS-CoV-2: Method Development and Validation. Bio-protocol 16(24): e5903. DOI: 10.21769/BioProtoc.5903.
Do you have any questions about this protocol?
Post your question to gather feedback from the community. We will also invite the authors of this article to respond.
Share
Bluesky
X
Copy link