Real-time deformability cytometry (RT-DC)

AU Alena Uvizl
RG Ruchi Goswami
SG Shanil Durgeshkumar Gandhi
MA Martina Augsburg
FB Frank Buchholz
JG Jochen Guck
JM Jörg Mansfeld
SG Salvatore Girardo
request Request a Protocol
ask Ask a question
Favorite

The RT-DC measurements were performed by using the same setup and device described in a previous publication.49 Briefly, the cells were trypsinised and were centrifuged at 100g for 5 min. The supernatant was removed, and the cells were resuspended at a concentration of around 5 × 106 cells per ml in 1× PBS buffer with 0.5% (w/v) methylcellulose having viscosity of 15 mPa s. The cell samples were passed through RT-DC devices with 30 μm square channel geometry at a flow rate of 0.16 μL s−1. All the samples were processed at room temperature within 30 min of preparation. All analysis was done using the open source software ShapeOut.50 The Young's moduli of the cells were calculated using ShapeOut after filtering out the events falling out of area range of 80–750 μm2 and area ratio range of 1–1.05.