2.13. The method of analysis sample using HPLC or HPLC-MS/MS

HC Hang Chen
SH Sifan Huang
HW Heyi Wang
XC Xinmei Chen
HZ Haiyan Zhang
YX Youfa Xu
WF Wei Fan
YP Yun Pan
QW Qiuyan Wen
ZL Zhizhe Lin
XW Xuena Wang
YG Yongwei Gu
BD Baoyue Ding
JC Jianming Chen
XW Xin Wu
ask Ask a question
Favorite

Using Agilent 1260II HPLC system for high performance liquid chromatography (HPLC) analysis(Agilent, USA). Samples were separated using an Agilent EclipsePlus C18 column (250 × 4.6 mm, 5 μm). The mobile phase was a mixture of methanol & water (95:5, v/v), and the flow rate was 1.0 mL/min. Absorbance was monitored at 227 nm.

Using Agilent Eclipse plus C18 column (4.6 mm × 250 mm, 5 μm), gradient elution was performed at a UV detection wavelength of 227 nm. The elution program was water(A): acetonitrile(B), 0-14 min: 68%A-32%B; 1 4∼27 min: 100%A-0%B; 2 7 ~28 min: 68%A-32%B, 2 9 ∼ 33 min and the flow rate of 1 mL/min, inject 20 μL of the plasma sample to be tested.

MS conditions are shown in the table below: