Negative-stain sample preparation, data collection, and initial model building

CX Cong Xu
YW Yanxing Wang
CL Caixuan Liu
CZ Chao Zhang
WH Wenyu Han
XH Xiaoyu Hong
YW Yifan Wang
QH Qin Hong
SW Shutian Wang
QZ Qiaoyu Zhao
YW Yalei Wang
YY Yong Yang
KC Kaijian Chen
WZ Wei Zheng
LK Liangliang Kong
FW Fangfang Wang
QZ Qinyu Zuo
ask Ask a question
Favorite

For the negative-stain sample, a volume of 5 μl of SARS-CoV-2 S-ACE2 sample was placed on a plasma-cleaned copper grid for 1 minute. Excess sample on the grid was blotted off using filter paper, and a volume of 5 μl of 0.75% uranyl formate (UF) (Sigma-Aldrich) was added to wash the grid. After blotting, another volume of 5 μl of 0.75% UF was placed on the grid again for 1 minute to stain. Grids were visualized under a Tecnai G2 Spirit 120-kV transmission electron microscope (Thermo Fisher Scientific), and micrographs were taken using an Eagle camera with a nominal magnification of 67,000 ×, yielding a pixel size of 1.74 Å. A total of 41,827 particles were autopicked in EMAN2 (48). After 2D classification, we selected good averages with 13,047 particles for initial model building, which were performed in Relion 3.0 (49).