Determination of MSN Size and Zeta Potential

LF Lara G. Freidus
PK Pradeep Kumar
TM Thashree Marimuthu
PP Priyamvada Pradeep
YC Yahya E. Choonara
request Request a Protocol
ask Ask a question
Favorite

Size and zeta potential analyses were performed on the Zetasizer Nano ZS (Malvern, United Kingdom.). Dilute 10 mL suspensions of nanoparticles were sonicated for 30 min at room temperature prior to analysis to reduce particle aggregation. 2 mL of the dispersed samples were filtered into disposable cuvettes and the size, polydispersity index (PDI) and zeta potential values were recorded at 25°C on the ZetaSizer Nano ZS.