2.3.2. Precision

YH Yanyan Hou
YC Yongquan Chen
JS Jieli Sun
JG Jinting Geng
HJ Hongwei Jin
ZZ Zhongying Zhang
request Request a Protocol
ask Ask a question
Favorite

According to the EP5-A guideline published by CLSI,[11] precision was evaluated by calculating the coefficients of variations (CVs) of intra-assay and inter-assay. The experiments were performed using 2 concentrations (low concentration and concentration) of serum samples, which were analyzed twenty times, respectively. CV (%) = (Standard deviation/average) 100%. When the CV was no more than 10.0%, the precision would be accepted.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

0/150

tip Tips for asking effective questions

+ Description

Write a detailed description. Include all information that will help others answer your question including experimental processes, conditions, and relevant images.

post Post a Question
0 Q&A