Qualitative examination for S. mansoni circulating cathodic antigens (CCA)

MC Mamoudou Cisse
IS Ibrahim Sangare
AD Arthur D. Djibougou
MT Marc C. Tahita
SG Souleymane Gnissi
JB Jonathan K. W. Bassinga
SK Salifou Konda
AD Abdoulaye H. Diallo
ask Ask a question
Favorite

All urine samples were tested for S. mansoni CCA using a commercially available point-of-care (POC-CCA) cassette test (batch numbers 191031120, ICT INTERNATIONAL, Noordhoek, South Africa) performed at room temperature according to the manufacturer’s instructions on the day of the sample collection. Briefly, 2 drops of urine were added to the well of the testing cassette and allowed to absorb. The test results were read visually 20 min later. In cases when the control bands did not develop, the test was considered invalid. Invalid tests were repeated using the same urine sample. Valid tests were scored as negative or positive. Trace (weak band) was also considered positive [30]. All the tests were read independently by two laboratory technicians, and in case of discordant results, the senior parasitologist examined it until an agreement was reached.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

post Post a Question
0 Q&A