For robust quantification, we curve-fitted pseudo-Voigt profiles to creatinine, myo-inositol, dimethyl sulfone, and valine NMR signals. This method allows determination of the goodness of fit by assessing differences between the spectrum and fitted profile, and thus indicates when the quantification is unreliable due to interference by other metabolites. For analytical validation of biomarker quantification, precision, linearity, and bias were analyzed.
For analytical validation of biomarker quantification, a 50-mL serum pool of 25 pediatric outpatients between 8 and 17 years of age with eGFRCrea < 90 mL/min/1.73 m2 at the day of serum collection was collected between October 2016 and January 2017 at Hannover Medical School (INPREM cohort). Sampling was ethically approved by Hannover Medical School’s institutional review board (No. 3396-2016, dated 15 September 2016). Individual serum samples were stored no longer than 2 h at room temperature before interim storage at −20 °C. Long-term storage was at −80 °C. Samples were pooled just before preparation for NMR analysis. Additional human serum pools were purchased by Bavarian Red Cross (Regensburg, Germany) and stored at −20 °C until utilization. Analytical precision was assessed using three distinct serum pools: the first two pools consisted of commercially available Bavarian Red Cross (adult normal GFR, adult low adult GFR) serum, while the third pool contained pooled sera collected from 25 pediatric patients enrolled in the INPREM study (pediatric low GFR, see above for details). Serum pools were measured in five runs with three replicates per measurement on a single NMR device, resulting in 15 NMR measurements for each pool. Within-run, between-run, and total variation were analyzed using a fully nested model II ANOVA, and analytical coefficients of variation (CV) were computed.
In the linearity study, a working range for each biomarker was determined within which the relationship between the observed values and the true concentrations of the metabolite of interest is linear. Therefore, a serum pool with high metabolite serum levels (i.e., high pool) was prepared by spike-in before preparing linearity samples. A linear dilution in 13 concentration steps down to 0% of the high pool was prepared. For each concentration step, three replicates were analyzed.
The bias analysis measured the closeness between NMR-measured metabolite concentrations and spike recovery or whenever available clinical chemistry reference values. For dimethyl sulfone, myo-inositol, and valine, bias was determined by spike recovery experiments. To generate the samples, non-modified and dialyzed sera were pooled to generate ‘mini-pools’. Additives spiked with the respective metabolites were generated for 12 different concentration levels. The respective controls consisted of the same mini-pool prepared with additives, but without the respective metabolite. In total, 138 mini-pools were used in the spike recovery experiment. For creatinine, standard clinical chemistry reference methods were used (Creatinine reagent OSR6678 on a Beckman Coulter AU640 analyzer, Beckman Coulter Inc., Brea, CA, USA). In total, 120 human serum mini-pools, partly spiked or dialyzed to cover a broader concentration range, were measured in duplicates by NMR and by the reference method. In total, 115 mini-pools were used in the study.
Do you have any questions about this protocol?
Post your question to gather feedback from the community. We will also invite the authors of this article to respond.