2.2. Defining an rRT-PCR CT-Value Range for Potential Infectious Samples

NK Niko Kohmer
TT Tuna Toptan
CP Christiane Pallas
OK Onur Karaca
AP Annika Pfeiffer
SW Sandra Westhaus
MW Marek Widera
AB Annemarie Berger
SH Sebastian Hoehl
MK Martin Kammel
SC Sandra Ciesek
HR Holger F. Rabenau
ask Ask a question
Favorite

Three quantitative comparison samples containing 105, 106, and 107 SARS-CoV-2 (BetaCoV/Munich/ChVir984/2020) RNA copies/mL were used to generate a standard curve and to calculate the viral RNA copies/mL (Table S1 and Figure S1). The comparison samples were provided by INSTAND e.V. In total, 10 aliquots of each suspension were tested on two different days (5 aliquots/day) to verify the intra- and inter-assay reproducibility.