Cryo-EM sample preparation and data acquisition

HS Hendrik R. Sikkema
MN Marco van den Noort
JR Jan Rheinberger
MB Marijn de Boer
SK Sabrina T. Krepel
GS Gea K. Schuurman-Wolters
CP Cristina Paulino
BP Bert Poolman
ask Ask a question
Favorite

Freshly prepared OpuA nanodiscs were concentrated to at least 1 mg/ml using a Vivaspin 500 (50 kDa cutoff) concentrator. Before freezing, OpuA was diluted to 1 mg/ml and respective additives were added (e.g., 100 μM glycine betaine, 10 μM cyclic-di-AMP, 10 μM MgAMP-PNP, or 10 mM MgATP). Holey carbon grids (Quantifoil; Au R1.2/1.3, 300 mesh) were glow-discharged for 20 s at 5 mA. Sample (2.8 μl) was applied on the grids, after which they were blotted for 3 to 5 s in a Vitrobot Mark IV (Thermo Fisher Scientific) at 15° to 22°C and 100% humidity and plunge-frozen into a liquid ethane/propane mixture. The grids were then stored in liquid nitrogen until further use.

Data collection was performed in-house on a 200-keV Talos Arctica microscope (Thermo Fisher Scientific) equipped with a K2 and a post-column BioQuantum energy filter (Gatan) operated in zero-loss mode, with a 20-eV slit, and a 100-μm objective aperture. Automatic collection was done at a calibrated magnification of 49,407× (1.012 Å pixel size) and a nominal defocus range from −0.8 to −1.9 μm using EPU (Thermo Fisher Scientific). Holes were selected with the help of an in-house–written script that calculates ice thickness within DigitalMicrograph (Gatan). Each movie consisted of 60 frames with a total exposure time of 9 s and a dose of 53 electrons/Å2 (0.883 e2 per frame). FOCUS software (38) was used for on-the-fly quality control, and settings were adjusted if necessary.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

post Post a Question
0 Q&A