Interfacial rheology measurements

ZC Zhu Cheng
CS Carolyn R. Shurer
SS Samuel Schmidt
VG Vivek K. Gupta
GC Grace Chuang
JS Jin Su
AW Amanda R. Watkins
AS Abhishek Shetty
JS Jason A. Spector
CH Chung-Yuen Hui
HR Heidi L. Reesink
MP Matthew J. Paszek
request Request a Protocol
ask Ask a question
Favorite

A standard rotational rheometer MCR 302 (Anton Paar) equipped with a Peltier temperature device (P-PTD 200/80I) was used. Temperature was set to 25°C for all measurements. A biconical disk–based measuring system (BiC68-5) in conjunction with an Interfacial Rheology System cell was used to measure the interfacial moduli. A normal force assisted surface detection methodology was used to position the bicone geometry accurately at the buffer-silicone oil interface. A special interfacial analysis software package based on the unique solution of the full flow field for a biconical geometry (49) was used for the calculation of the absolute interfacial rheological properties.