request Request a Protocol
ask Ask a question
Favorite

A central venous catheter (Arrow, Asheboro, NC, USA) was inserted in the subclavian vein, and was confirmed to be placed in the superior vena cava by bedside echo exam for analysis. The zero point was corrected and the CVP read. The PiCCO catheter (4F, PULSION, Feldkirchen, Germany) was placed via the femoral artery prior to connecting the monitor. The temperature probe was connected to the subclavian vein catheter. After the monitor was ready, mean arterial pressure (MAP), CI, GEDVI, and ITBVI were recorded. During measurement, 20 ml of saline in 4°C was injected quickly (within 5 s), measuring at least 3 times in a row, and using measurements whose variation was less than 15%.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

0/150

tip Tips for asking effective questions

+ Description

Write a detailed description. Include all information that will help others answer your question including experimental processes, conditions, and relevant images.

post Post a Question
0 Q&A