LH and FSH levels

VP Valentina Rodriguez Paris
SS Samantha M. Solon-Biet
AS Alistair M. Senior
ME Melissa C. Edwards
RD Reena Desai
NT Nicodemus Tedla
MC Madeleine J. Cox
WL William L. Ledger
RG Robert B. Gilchrist
SS Stephen J. Simpson
DH David J. Handelsman
KW Kirsty A. Walters
request Request a Protocol
ask Ask a question
Favorite

Blood was collected from females by cardiac exsanguination under ketamine/xylazine anesthesia, and collected serum was stored at −20 °C. Mice were collected at the diestrus stage for cycling mice and pseudo-diestrus for acyclic PCOS mice. FSH and LH levels in 10 µl of sera diluted in 15 µl serum dilution buffer (1 in 2.5 dilution) was measured using EMD Millipore’s MILLIPLEX® MAP 96-well Mouse Pituitary Magnetic Bead custom Kit (Cat# MPTMAG-49K) according to manufacturer’s instructions (EMD Millipore, Billerica, MA, USA). Luminex MAGPIX instrument (Luminex Corporation, Northbrook, IL, USA) calibrated with MAGPIX Calibration and Performance Verification Kits (Millipore) with xPONENT software (Luminex) was used to acquire data. Saved data was then analyzed as the Median Fluorescent Intensity (MFI) using spline curve-fitting for calculating analyte concentrations in samples with Multiplex Analyst software Version 5.1 (Luminex).

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

post Post a Question
0 Q&A