The high-resolution mass spectrometer used was an Orbitrap Q-Exactive (Thermo Scientific, Milan, Italy) equipped with an HESI II interface set in positive (ESI(+)) polarity. A Q-Exactive mass spectrometer was operated in full scan (70 000 FWHM). Mass spectra were recorded from m/z 400 to 2000. Acquisition parameters were optimized by direct infusion of standard working solutions of the analytes (50 µM in water) at a flow rate of 3 μL min−1 into the ESI interface of the mass spectrometer. The parameters were optimized as follows: spray voltage set at 3.5 kV (ESI(+)), capillary temperature set at 300 °C; sheath, auxiliary, and sweep gases (nitrogen) fixed at 5, 0, and 0 (arbitrary unit), respectively. Acquisition data were recorded and elaborated using Xcalibur™ version 3.0 software from Thermo Fisher.
Do you have any questions about this protocol?
Post your question to gather feedback from the community. We will also invite the authors of this article to respond.