Functional slit-lamp biomicroscope (FSLB) for measuring conjunctival microcirculation

CS Ce Shi
HJ Hong Jiang
GG Giovana Rosa Gameiro
JW Jianhua Wang
request Request a Protocol
ask Ask a question
Favorite

FSLB settings and image procedures have been well described in previous studies [2, 8], and its repeatability had been validated [7]. In brief, the FSLB was modified from a traditional slit-lamp by adding a digital camera which has a special function called Movie Crop Function (MCF). The MCF enables the addition of a 7× magnification, which can be combined with the slit-lamp magnification (30×), resulting in extremely high magnification (~ 210×) for imaging the motion of the cluster of red blood cells (Fig. (Fig.1).1). In this study, the field of view of the FSLB was 0.9 × 0.7 mm. The pixel interval on the acquired video was 1.4 μm. The other settings of the camera were the same as in previous studies [7, 9]. To measure the mean BFV and blood flow rate (BFR), six different locations approximately 1 mm away from the limbus were recorded. The measurement was taken on conjunctival venules because the majority of the conjunctival vessels are venules. Custom software was used to obtain BFV and BFR and the detailed image processing procedures have been reported previously [7, 9].

Blood flow velocity and blood flow rate of the conjunctiva and retina. FSLB and RFI were used to measure the blood flow velocity and flow rate in the conjunctiva (a) and retina (b). A total of six fields of view (red rectangles inside the white rectangle) were imaged using FSLB in the temporal side of the bulbar conjunctiva 1 mm apart from the limbus. With the extremely high magnification (~ 210 ×), the cluster of red blood cells was visualized (red arrowhead in the insert), which facilitated the measurement of blood flow velocity and flow rate. Retinal blood flow velocity was measured in the 2nd and 3rd branches of retinal arterioles (red) and venules (purple) which were overlaid with the measured blood flow velocity (note: negative values indicate arteriolar velocity). To estimate the blood flow in the retina, a circle (white, diameter = 2.5 mm) centered on the fovea was drawn. Vessel diameters of the vessels crossing the circle were measured in the arterioles (red dots) and venules (yellow dots). The velocity and vessel diameter were used to calculate the flow rate

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

0/150

tip Tips for asking effective questions

+ Description

Write a detailed description. Include all information that will help others answer your question including experimental processes, conditions, and relevant images.

post Post a Question
0 Q&A