2.4. SERS Measurement

MC Mingming Cheng
CL Chenyan Li
WL Weijun Li
YL Yingkai Liu
ask Ask a question
Favorite

First, rhodamine 6G was dissolved in purified water to obtain R6G solutions with different concentrations. By adding purified water to the cadmium nitrate solution, a 1.0 × 10−3–1.0 × 10−7 M solution was prepared as the probe molecule. After that, 5 μL of R6G solution (or cadmium nitrate solution) with different concentrations were dropwise dispersed on the above-mentioned silicon wafer (0.6 cm × 0.6 cm) substrate containing Ag@CuO FM and then naturally dried. The detection of SERS was carried out under a 50× objective lens and confocal Raman system with excitation wavelengths of 532 nm and 0.785 μm2, respectively. Raman grating with a 1200 L/mm, the spectral data acquisition 3 times, each acquisition time is 10 s.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

post Post a Question
0 Q&A