2.1. Laboratory Measurements

PV Peeradon Vibhatavata
PP Pongthep Pisarnturakit
MB Mongkol Boonsripitayanon
PP Paveena Pithuksurachai
NP Nattachet Plengvidhya
SS Sirinart Sirinvaravong
ask Ask a question
Favorite

Serum 25(OH)D and iPTH levels were measured using an electrochemiluminescence immunoassay (ECLIA) using a cobas® 8000 modular analyzer (Hitachi High-Technologies Corporation, Tokyo, Japan), and serum total calcium, phosphate, albumin, and magnesium levels were measured by standard methods using a cobas® 8000 modular analyzer (Hitachi High-Technologies Corporation) in the Department of Clinical Pathology, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand. The reference ranges of laboratory parameters were as follows: 8.8–10.2 mg/dL for serum calcium; 2.5–4.5 mg/dL for serum phosphate; ≥30–100 ng/mL for serum 25(OH)D; and 10–65 pg/mL for serum iPTH. The intra-assay and interassay coefficients of variation (CVs) of these tests were 0.7% and 0.8% for serum calcium; 0.8% and 1.4% for serum phosphate; 3.3% and 4.5% for serum 25(OH)D; and 0.9% and 1.5% for serum iPTH, respectively.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

post Post a Question