Single-Particle Cryo-EM sample vitrification

YL Yang Liu
KZ Keda Zhou
NZ Naifu Zhang
HW Hui Wei
YT Yong Zi Tan
ZZ Zhening Zhang
BC Bridget Carragher
CP Clinton S. Potter
SD Sheena D'Arcy
KL Karolin Luger
request Request a Protocol
ask Ask a question
Favorite

Reconstituted complexes were concentrated to 7-10 μM by Amicon Ultra −4 centrifugal filter (Ultracel 50K, Millipore). Fresh detergent (CHAPS, 0.25%-0.5%) was added before grid preparation. In order to minimize sample interaction with the air-water interface and to keep the complex intact29, the complex was vitrified using a commercial prototype (from TTP Labtech) of the Spotiton robot14,30,31. The time between sample being spotted onto the nanowire grid32 and vitrification into liquid ethane was 133 ms. Grids were plasma-cleaned (Gatan Solarus) for 5 s using an H2/O2 mixture prior to vitrification. The robot chamber was operated at room temperature and ~80% humidity in the sample and plunging area.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

post Post a Question
0 Q&A