Micro-CT imaging

BC Barden Chan
MC Michelle Clasquin
GS Gromoslaw A. Smolen
GH Gavin Histen
JP Josh Powe
YC Yue Chen
ZL Zhizhong Lin
CL Chenming Lu
YL Yan Liu
YC Yong Cang
ZY Zhonghua Yan
YX Yuanfeng Xia
RT Ryan Thompson
CS Chris Singleton
MD Marion Dorsch
LS Lee Silverman
SS Shin-San Michael Su
HF Hudson H. Freeze
SJ Shengfang Jin
request Request a Protocol
ask Ask a question
Favorite

Mice were anesthetized by 1% pentobarbital sodium (50 mg/kg, i.p.) for micro-CT scans with a GE eXplore Locus (GE Healthcare). All scans were performed in the same mode: 80 kV X-ray voltage, 450 μA current, 0.045 mm effective pixel size, 360° tomographic rotation and a rotation step of 0.9° with an exposure time of 400 ms for each view. All scans were reconstructed as 3D images by eXplore MicroView 2.0 analysis software with the same parameters.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

0/150

tip Tips for asking effective questions

+ Description

Write a detailed description. Include all information that will help others answer your question including experimental processes, conditions, and relevant images.

post Post a Question
0 Q&A