Purified phage particles at a concentration of 109 PFU/mL were deposited on 300 mesh carbon-coated copper grids (Agar Scientific, United Kingdom) and negatively stained with 1% uranyl acetate (pH 4) as follows: the surface of the copper grids were ionized for 2 min immediately prior to sample deposition. 5 μL of phage lysate were spotted on the surface of the grid and allowed to stand for 1 min. This was followed by 30 sec negative staining and subsequent air drying of the grid. Visualization was performed using a JEOL JEM-1400Plus TEM, operated at 120 kV (pixel size = 0.1 nm), equipped with a Gatan OneView 4K camera with automatic drift correction.