LC-MS analysis of procainamide-labeled N-glycans

SC Silvia Crescioli
GC Giulia Chiaruttini
SM Silvia Mele
KI Kristina M. Ilieva
GP Giulia Pellizzari
DS Daniel I.R. Spencer
RG Richard A. Gardner
KL Katie E. Lacy
JS James F. Spicer
AT Andrew N.J. Tutt
GW Gerd K. Wagner
SK Sophia N. Karagiannis
ask Ask a question
Favorite

Samples were analyzed by using HILIC-LC on an Ultimate 3000 UHPLC with a BEH-Glycan 1.7 μm, 2.1 × 150 mm column (Waters, Milford, Mass) at 40°C with a fluorescence detector (λex = 310 nm, λem = 370 nm) controlled by Bruker HyStar 3.2 (Bruker, Billerica, Mass). Buffer A was 50 mmol/L ammonium formate made from LudgerSep N Buffer stock solution, pH 4.4 (LS-N-BUFFX40); buffer B was acetonitrile (acetonitrile 190 far UV/gradient quality; Romil #H049). Samples were injected in 25% aqueous/75% acetonitrile; injection volume was 25 μL. Separation was performed with a linear gradient of 76-53% ACN at 0.4 mL/min in a 71-minute analytic run. PROC-labeled glucose homopolymer was used as a system suitability standard, as well as an external calibration standard, for glucose unit allocation for the system. Chromeleon Data Software (version 7.2; Thermo Fisher Scientific, Waltham, Mass) with a cubic spline fit was used to allocate glucose unit values to peaks. A Bruker amaZon Speed ETD electrospray mass spectrometer (Bruker) was coupled directly after the ultra-HPLC with fluorescence detection without splitting. The instrument scanned samples in maximum resolution mode with a positive ion setting, mass spectrometry (MS) plus 3 MS/MS scans, nebulizer pressure of 14.5 psi, nitrogen flow of 10 L/min, and capillary voltage of 4500 V. MS/MS was performed on 3 ions in each scan sweep with a mixing time of 40 ms. MS data were analyzed by using Bruker Compass DataAnalysis 4.1 software.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

post Post a Question
0 Q&A