Phenazine Analysis

TA Theresia D. Askitosari
SB Santiago T. Boto
LB Lars M. Blank
MR Miriam A. Rosenbaum
request Request a Protocol
ask Ask a question
Favorite

The quantification of phenazines was performed using high-performance liquid chromatography (HPLC) with a C18 column (Waters Corporation) equipped with a photodiode array UV/VIS detector (LC-168, Beckmann), which detected and quantified the PCA at 366 nm and PYO at 280 nm. Separation was achieved with a gradient of 0.1% trifluoroacetic acid (TFA) in acetonitrile (solution A) and 0.1% TFA in water (solution B) as mobile phase (with solution A at 15% for 2 min, 100% for 15 min, and 5% for 3 min) at a flow rate of 0.8 mL/min and a column temperature of 20°C (Kern and Newman, 2014). Analytical grade PCA (Princeton BioMolecular) and PYO (Cayman Chemical) were used as standard solutions of phenazines.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

post Post a Question
0 Q&A