HR MRI and MRA examination

JL Jung Eun Lee
YL Yong Jae Lee
SC Soo Yoon Chung
HC Hee Woo Cho
BP Byoung Jin Park
DJ Dong Hyuk Jung
request Request a Protocol
ask Ask a question
Favorite

All MRI examinations were performed on a 3.0 T MRI system (Signa 3.0T, General Electric, Milwaukee, WI, USA; Achieva 3.0T, Philips Medical Systems, Best, Netherlands). Diffusion-weighted images were obtained using the following parameters: TR/TE 9000/81.4 ms, matrix 128× 128, FOV 240×240 mm, slice thickness 4 mm, b = 1000 s/mm2. The voxel volumes were 14 mm3. Three-dimensional brain MRA was performed for arterial and venous phase imaging with following parameters: repetition time 4900 ms, echo time minimum, a 30×27 cm field of view, image matrix 256×192, flip angle 20°, and section thickness 14 mm. The dose of gadobutrol (Gadovist®; Schering, Berlin, Germany) was 0.2 mL/kg of body weight. The contrast agent injected with a power injector (Spectris MR Injector®; Medrad, Pennsylvania, USA) at a rate of 2 mL/sec. When the contrast was visually detected in the carotid artery by the technologist on axial plane during injection of the contrast, the coronal 3D gradient echo sequence was initiated; and image acquisition time 2–3 minutes. After acquisition of arterial phase of MRA, venous phase images were obtained. Major brain and neck artery atherosclerosis was assessed on MRA using picture archiving communication system technology (PACS; GE Medical Systems, Milwaukee, WI, USA).