MitoXpress® Xtra–Oxygen Consumption Assay (Luxel Biosciences) was used for the direct, real-time analysis of cellular respiration and mitochondrial function. MitoXpress® Xtra is quenched by O2 through molecular collision, and thus the amount of fluorescence signal is inversely proportional to the amount of extracellular O2 in the sample. HUVEC were serum starved for 1 hour and incubated with 3% plasma for 4 hrs. After incubation cells were washed with clear respiration media. The MitoXpress-Xtra-HS probe was added to cells in accordance with manufacturer’s instructions. Oxygen consumption was measured using time-resolved fluorescence (TR-F) with a dual delay of 30 μs and 70 μs using a VarioSkan fluorescence plate reader (Thermo-Scientific). Rate of oxygen consumption was determined from the slope of fluorescence vs. time for each sample using relative fluorescence units/hour.