Error analyses in CBF and V0 calculation

MI Muhammad M. Islam
TT Tetsuya Tsujikawa
TM Tetsuya Mori
YK Yasushi Kiyono
HO Hidehiko Okazawa
ask Ask a question
Favorite

Error analyses were performed to investigate the biases of the CBF and V 0 estimations with simulated delay time shifts. Several small ROIs 10-mm in diameter, including about 30 voxels, were drawn on the brain cortex of parametric images for various values of CBF (30–70 mL/min/100 g) and V 0 (0.3–2.0 mL/100 g) to obtain tissue TACs from dynamic PET data. The arterial input function after the delay and dispersion correction for the brain tissue TAC (=C a) was shifted from −3 to 3 s in steps of 0.5 s in the analyses. Using these two TACs of tissue activity and shifted C a, CBF and V 0 were recalculated based on a 1-TCM. Percent changes from the original CBF and V 0 values were calculated and plotted to observe the effect of delay estimation errors on parametric values.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

post Post a Question
0 Q&A