A radiochromic film (Gafchromic EBT3-1417, Lot # 09241801, Ashland Global Specialty Chemicals Inc., DE, USA) was roughly cut into small rectangular pieces and placed at 5 mm intervals in a 3D-printed holder that had 15 slots 5 mm apart. The holder with pieces of the radiochromic film was placed vertically under the X-ray beam with the radiochromic piece in the 0 mm slot directly in contact with the focusing optics (polycapillary) of the X-ray and exposed to the X-ray beam (at 50 kV and 600 μA) for 30 s. The spot size was measured with ImageJ software.
Do you have any questions about this protocol?
Post your question to gather feedback from the community. We will also invite the authors of this article to respond.
Tips for asking effective questions
+ Description
Write a detailed description. Include all information that will help others answer your question including experimental processes, conditions, and relevant images.