High-resolution mass spectrometry (HR-MS) measurements for synthesized compounds were performed using a Q Exactive mass spectrometer (Thermo Fisher Scientific, Waltham, MA) with an electrospray ionization (ESI) probe. Full scan mass spectra were recorded in positive mode. The conditions for ESI were as follows: sheath gas flow rate, 5 units; capillary temperature, 320 °C; spray voltage, 3.5 kV; probe heater temperature, 350 °C; S-lens RF level, 50 units. Products were dissolved in a 1:1 (v/v) mixture of water–acetonitrile containing 0.1% formic acid at a concentration of 1 μg/mL and infused into the spectrometer at a flow rate of 10 μL/min.
Copyright and License information: The Author(s) ©2021 Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.