PWV was measured using a volume-plethysmograph (PWV/ABI; Omron Healthcare Co. Ltd., Japan). A detailed description of this device and the validity and reproducibility of its measurements have been published elsewhere16. Brachial-to-ankle PWV (baPWV) was calculated using the time interval between the wave fronts of the brachial and ankle waveforms (∆Tba) and the path length from the brachium to the ankle. Path lengths from the suprasternal notch to the brachium (Lb) or ankle (La) were calculated using the formulae: Lb = 0.2195 × height + 2.0734; La = 0.8129 × height + 12.328. Then, baPWV was calculated using the equation (La−Lb)/∆Tba.