2.3.1. UV–Visible Diffuse Reflectance Spectroscopy (UV–Vis DRS)

HM Hanieh Mousavi
TS Thomas D. Small
SS Shailendra K. Sharma
VG Vladimir B. Golovko
CS Cameron J. Shearer
GM Gregory F. Metha
ask Ask a question
Favorite

UV–Vis DRS measurements were used to infer the extent of Au101NC agglomeration and to confirm that co-catalysts (rGO and Au101NC) did not affect the bandgap of the photocatalyst (AlSrTiO3). UV–Vis DRS measurements were obtained using a spectrophotometer (Cary 5000 UV–Vis–NIR) fitted with a Praying Mantis Diffuse Reflection accessory (Harrick, DRP-SAP). A PTFE disc was used as reflectance standard. For each measurement, the sample holder was filled with approximately 15 mg of the solid-state sample and the reflectance was measured from 200–800 nm.