Coixol HPLC quantitative precision test

XW Xing-Chen Wang
XS Xin-Yu Shen
LC Lin Chen
RW Rong Wei
MW Ming-Yuan Wei
CG Cai-Hong Gu
RX Rong-Rong Xu
SD Sheng-Qing Ding
BP Bo Pan
request Request a Protocol
ask Ask a question
Favorite

Standard coixol solutions (10, 50, and 150 μg/mL) were analysed separately under established chromatographic conditions. The results are presented in Table 1. The relative standard deviation (RSD%) of the peak area at different concentrations was 0.74%, 0.99%, and 0.77%, respectively. These RSD values indicate satisfactory precision in the detection method.

The accuracy test of coixol.