Determination of Sterilization Dose according to ISO 11137

AS Agus Susanto
IK Ira Komara
MB Maria Theresia Beatrix
FL Fajar Lukitowati
AA Amaliya Amaliya
IH Ina Hendiani
AM Aldilla Miranda
ask Ask a question
Favorite

The estimated radiation dose produced a sterility assurance level (SAL) of 10 −2 , and then, the sterilization dose was determined based on the table on ISO 11137-2 ( Table 1 ) and based on the initial number of microorganism populations from the bioburden test, and we determined doses based on SAL 10 −6 . 24

Abbreviation: CFU, colony forming unit.

Source: Table copied from Table 5. ISO 11137–2 by retrieving only part of the data.