Our SRS imaging system consists of a dual output femtosecond (80 MHz repetition rate) laser (InSight DS+, MKS) with tunable (680 to 1300 nm) and fixed (1040 nm) outputs, a delay line to attain and adjust beam overlap, and a custom-built laser scanning microscope.26 Spectral focusing is achieved using compact adjustable-dispersion TIH53 glass blocks27 to chirp the femtosecond pulses into the picosecond domain, at 2.06 ps and 1.1 ps for pump and Stokes, respectively, and adjusting pulse overlap with the delay line, allowing for an spectral resolution. We measured the stimulated Raman loss (SRL) signal using lock-in detection with the Stokes beam modulated by an acousto-optic modulator (1205C-2, Isomet) at 4.9 MHz. The pump beam intensity was detected by a photodiode (FDS1010, Thorlabs) biased to 60 V, filtered through a 2 to 7 MHz electronic bandpass filter, and delivered to a lock-in amplifier (MFLI, Zurich instruments) to extract the SRL signal. A pixel dwell time of was used for the SRS image. ScanImage28 (Version 5.6, Vidrio Technologies) was used for control of the laser scanning and image acquisition.
Do you have any questions about this protocol?
Post your question to gather feedback from the community. We will also invite the authors of this article to respond.