2.2.7. Differential Scanning Calorimetry (DSC)

SM Saima Mahmood
NK Nauman Rahim Khan
GR Ghulam Razaque
SS Shefaat Ullah Shah
MS Memuna Ghafoor Shahid
HA Hassan A. Albarqi
AA Abdulsalam A. Alqahtani
AA Ali Alasiri
HB Hafiz Muhammad Basit
ask Ask a question
Favorite

The films’ thermal characteristics were assessed using differential scanning calorimetry (DSC; PerkinElmer Thermal Analysis, USA) [89]. A 5–7 mg film sample was taken in the sample pan. The reference pan was left empty. Melting transition temperatures (T) of various films were recorded while continuously purging with N2 gas with a 40 mL/min flow rate and a heating scan rate of 10 °C/min from 0–400 °C. Each film peak’s transition temperature and enthalpy (∆H) values were calculated in triplicate, and the results were averaged.