2.2.1. Basic Qualification Tests

ZM Zahra Mansour
VU Verena Uslar
DW Dirk Weyhe
DH Danilo Hollosi
NS Nils Strodthoff
ask Ask a question
Favorite

To assess the reliability of the SonicGuard sensor across various instances, a total of six SonicGuard sensors were individually tested. Each sensor was placed at a distance of 1 meter from the sound source within an isolated room with free field conditions. A standardized reference signal, encompassing the frequency range of 1–20 kHz, was then employed. The performance of the SonicGuard sensors was subsequently compared to the calibrated microphone (MIC 255-1) to gauge their accuracy.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

post Post a Question
0 Q&A