Open circuit spirometry (K4b2; COSMED USA) was used to provide breath-by-breath cardiopulmonary data during all ADL activities, using methods previously described.20 In brief, the device was warmed up for 45 minutes and calibrated using a 3 liter syringe prior to testing. Following this, the O2 and CO2 analyzers were calibrated using a reference gas of known concentrations. Testing was initiated within 10 minute of completing calibration of the metabolic system. A snug fitting face mask was placed on participants and they underwent a period of acclimatization with the portable gear prior to performing the designated activities. Each participant’s mean oxygen consumption (VO2) for the eight activity types was determined and converted to Metabolic Equivalents of Task (METs; 1 MET=3.5 ml/kg/min). In addition, continuous heart rate monitoring provided a mean steady-state heart rate for each test.