cIEF was conducted according to the SCIEX application protocol. All cIEF experiments were performed on SCIEX PA800 plus instrument with a 50 µm i.d. neutral coated capillary (SCIEX P/N 477441) at the length of 30.2 cm. The UV detector was used to detect absorbance at 280 nm wavelength. A cIEF master mix solution was composed of 3 M urea-cIEF gel solution, pharmalyte 3–10 carrier ampholytes, cathodic stabilizer (500 mM arginine), anodic stabilizer (200 mM iminodiacetic acid), and five pI markers (10.0, 9.5, 7.0, 5.5, 4.1). Analytes (5 mg/mL) were also mixed with 10 µL of the master mix solution. All experiments were performed in triplicate. The pI values of the sample were calculated using qualitative analysis of a 32 Karat software.