4.7. Capillary Isoelectric Focusing (cIEF) Analysis

HC Heesu Chae
SC Seulki Cho
MJ Munsik Jeong
KK Kiyoung Kwon
DC Dongwook Choi
JL Jaeyoung Lee
WN Woosuk Nam
JH Jisu Hong
JL Jiwoo Lee
SY Seonjoo Yoon
HH Hyojeong Hong
request Request a Protocol
ask Ask a question
Favorite

cIEF was conducted according to the SCIEX application protocol. All cIEF experiments were performed on SCIEX PA800 plus instrument with a 50 µm i.d. neutral coated capillary (SCIEX P/N 477441) at the length of 30.2 cm. The UV detector was used to detect absorbance at 280 nm wavelength. A cIEF master mix solution was composed of 3 M urea-cIEF gel solution, pharmalyte 3–10 carrier ampholytes, cathodic stabilizer (500 mM arginine), anodic stabilizer (200 mM iminodiacetic acid), and five pI markers (10.0, 9.5, 7.0, 5.5, 4.1). Analytes (5 mg/mL) were also mixed with 10 µL of the master mix solution. All experiments were performed in triplicate. The pI values of the sample were calculated using qualitative analysis of a 32 Karat software.