4.4. Peptide Analysis by High-Performance Liquid Chromatography (HPLC)

EK Ekaterina P. Karaulova
EY Evgeny V. Yakush
TS Tatiana N. Slutskaya
LS Lidiya V. Shulgina
request Request a Protocol
ask Ask a question
Favorite

The molecular mass distributions of extracts were performed in a modular Agilent Technologies liquid chromatograph (Agilent Technologies 1260 Infinity, USA, CA) with a UV detection at 280 nm. All samples were prepared twice and measured by duplicate. A TSK gel G 3000PWXL column, 7.8 mm I.D. × 30 cm (TOSOH Corporation, Tokyo, Japan) at a flow-rate of 0.3 mL/min and a temperature of 25 °C were used. The mobile phase consisted of 0.1 N NaCI 20 mM Tris-HCI buffer, pH 7.8. The approximate molecular weight (MW) was determined using standard protein samples (Sigma-Aldrich Co., USA, MO) as reference: bovine serum albumin (MW 66.3 kDa), egg albumin (MW 44.3 kDa), myoglobin (MW 18.0 kDa) cytochrome C (MW 12.4 kDa), aprotinin (MW 6.5 kDa), bacitracin (MW 1.4 kDa). All samples were filtered through 0.2 µm syringe filter (Whatman, PVDF) before injection. The MW of peptides was calculated by the elution time.

Do you have any questions about this protocol?

Post your question to gather feedback from the community. We will also invite the authors of this article to respond.

0/150

tip Tips for asking effective questions

+ Description

Write a detailed description. Include all information that will help others answer your question including experimental processes, conditions, and relevant images.

post Post a Question
0 Q&A