(*contributed equally to this work) 发布: 2026年12月05日第16卷第23期 DOI: 10.21769/BioProtoc.5907 浏览次数: 31
评审: Anonymous reviewer(s)
Abstract
During oxidative phosphorylation, the mitochondrial membrane potential (ΔΨm) is generated by proton translocation across the inner mitochondrial membrane by respiratory chain complexes (I, III, and IV). This is a fundamental component of the mitochondrial proton-motive force that contributes to ATP synthesis and other mitochondrial bioenergetic functions. Melanoma, one of the most aggressive forms of skin cancer, is characterized by a significant metabolic reprogramming that contributes to disease progression, metastatic dissemination, and therapeutic resistance, highlighting the need for standardized and consistent approaches to assess mitochondrial function in living cancer cells. Mitochondrial function can be evaluated using biochemical assays, extracellular flux analysis, flow cytometry, or fluorescence microscopy. Among these approaches, live-cell fluorescence imaging enables quantitative measurement of ΔΨm while preserving cellular structure and single-cell resolution. In this protocol, we present a standardized workflow for measuring ΔΨm in living melanoma cells using tetramethylrhodamine ethyl ester (TMRE), normalizing TMRE fluorescence to mitochondrial content using the TMRE/MitoTrackerTM Green (MTG) fluorescence ratio, and using FCCP as a depolarization control. The workflow combines experimental optimization, standardized live-cell confocal imaging, and automated image analysis using a custom Fiji macro to support reliable fluorescence quantification. This protocol provides a robust, reproducible approach to assessing changes in mitochondrial membrane polarization in response to pharmacological, metabolic, or genetic perturbations.
Key features
• Quantitative measurement of mitochondrial membrane potential (ΔΨm) in living melanoma cells using TMRE fluorescence and normalization for mitochondrial content with MitoTrackerTM Green.
• Optimization of fluorescent probe concentrations, FCCP-mediated mitochondrial depolarization, and confocal acquisition parameters to achieve robust TMRE responses while preserving mitochondrial integrity for experimental reproducibility.
• Automated processing and quantification of three-dimensional fluorescence datasets using a custom-coded Fiji macro for z-projection, MitoTrackerTM Green-based mitochondrial segmentation, and region of interest (ROI) generation.
• Assessment of mitochondrial functional alterations induced by genetic, metabolic, or pharmacological perturbation in in vitro cell models.
Keywords: Mitochondrial membrane potentialGraphical overview
Workflow for live-cell confocal imaging and quantitative analysis of mitochondrial membrane potential using TMRE and MitoTracker Green
Background
Mitochondria are essential regulators of biosynthesis, cellular homeostasis, calcium signaling, energy production, and programmed cell death. Indeed, a key indicator of mitochondrial function is the mitochondrial membrane potential (ΔΨm), the electrical component of the proton electrochemical gradient produced across the inner mitochondrial membrane by proton pumping through the electron transport chain during oxidative phosphorylation [1–4]. Besides its role in ATP production, ΔΨm regulates mitochondrial protein import, calcium buffering, reactive oxygen species (ROS) production, metabolite transport, and mitochondrial integrity. Alterations in transmembrane potential are key indicators of mitochondrial dysfunction and are closely linked to cellular stress responses and metabolic reprogramming [3–5].
Dysregulation of ΔΨm has been associated with numerous pathological conditions, including aging, cardiovascular disorders, cancer, metabolic syndromes, and neurodegenerative diseases [6–10]. Beyond generating ATP, mitochondria in cancer pathophysiology also contribute to metabolic adaptation. For example, advanced melanoma is an aggressive malignancy characterized by substantial cellular plasticity that supports tumor progression, therapy resistance, and metastatic dissemination [11]. These phenotypic adaptations are sustained by mitochondrial remodeling, immune evasion, and metabolic reprogramming, processes that are associated with alterations in oxidative phosphorylation and cell fate regulation [11–14]. ΔΨm provides an integrated functional assessment of mitochondrial activity and metabolic adaptation [3–5,15].
Several methods can be used to analyze mitochondrial function [16,17]. High-resolution respirometry and extracellular flux analysis, including Seahorse-based assays, quantitatively measure oxygen consumption and provide functional information on mitochondrial respiration and electron transport chain activity. However, these approaches require specialized instrumentation and can be costly [16,18]. In addition, assay-specific changes in culture conditions, including altered serum or CO2 and repeated mixing cycles, may influence cellular bioenergetics and induce metabolic adaptations. Furthermore, biochemical assays measuring respiratory chain enzyme activity or ATP production provide useful functional measurements but often require cell lysis, precluding their application to living cells or tissues [16,19]. Using membrane potential–sensitive dyes such as tetramethylrhodamine ethyl ester (TMRE) or tetramethylrhodamine methyl ester (TMRM), flow cytometry enables rapid quantitative analysis of mitochondrial membrane potential across large cell populations but does not preserve spatial information on mitochondrial organization or cellular morphology [16,20–22]. In contrast, live-cell fluorescence microscopy enables assessment of mitochondrial membrane potential at the single-cell level while preserving spatial information on mitochondrial distribution, organization, and cellular morphology [23,24]. Among fluorescent approaches, TMRE is a well-established probe for assessing ΔΨm in living cells, as its accumulation within polarized mitochondria is driven by the negative membrane potential across the inner mitochondrial membrane [15,16,22,24]. A reduction of ΔΨm decreases TMRE accumulation, providing a sensitive fluorescence-based measure of mitochondrial depolarization. However, interpretation of TMRE fluorescence alone can be challenging, as variations in signal intensity may reflect differences in mitochondrial content, morphology, or cellular viability in addition to changes in ΔΨm. Confocal imaging parameters, image analysis, and variations in staining conditions can also affect quantitative fluorescence measurements and limit experimental reproducibility [24].
To strengthen the reliability of the TMRE-based measurements, this protocol incorporates the MitoTrackerTM Green (MTG) fluorescent dye and carbonyl cyanide p-trifluoromethoxy phenylhydrazone (FCCP). MTG is commonly used as a membrane potential–independent marker of mitochondrial content, allowing normalization of membrane potential–sensitive mitochondrial fluorescence [15,23]. Therefore, the TMRE/MTG ratio can be used as a relative measure of mitochondrial membrane potential normalized to mitochondrial content [22,23,25]. However, MTG fluorescence has occasionally been reported to be affected by membrane potential–altering treatments under specific experimental conditions. This should therefore be verified for the cell type and treatment under investigation (see General note 9). FCCP, a potent lipophilic mitochondrial uncoupler, induces mitochondrial depolarization and a concentration-dependent decrease in TMRE fluorescence. It is used as a positive control to confirm that the assay reliably detects modulations in ΔΨm [14,22,26,27]. This protocol integrates standardized live-cell confocal imaging and automated image analysis using a custom Fiji macro to improve experimental reproducibility and limit user bias. Although demonstrated here in a melanoma cell line, this approach is adaptable to other cell types and has previously been applied to KP-4 pancreatic cancer cells to investigate bioenergetic responses and mitochondrial remodeling following experimental perturbations [23,24].
Materials and reagents
Biological materials
1. MeWo human melanoma cell line, NF1-mutant (RRID: CVCL_0445), was a generous gift from Ian Watson (McGill University)
Reagents
1. Tetramethylrhodamine ethyl ester (TMRE) (Thermo Fisher Scientific, catalog number: T669)
2. MitoTrackerTM Green FM (Thermo Fisher Scientific, catalog number: M7514)
3. Carbonyl cyanide p-trifluoromethoxyphenylhydrazone (FCCP) (Cayman Chemical, catalog number: 15218)
4. RPMI 1640 medium (Wisent, catalog number: 350-000)
5. RPMI 1640 medium with L-glutamine, without phenol red (Wisent, catalog number: 350-046)
6. Fetal bovine serum (FBS), premium (Wisent, catalog number: 090-150)
7. Penicillin–streptomycin solution 100×, sterile (Wisent, catalog number: 450-201)
8. Dimethyl sulfoxide (DMSO) (Fisher Scientific, catalog number: BP231100)
Solutions
1. Complete RPMI 1640 cell culture medium (see Recipes)
2. Phenol red-free imaging medium (see Recipes)
3. FCCP working solution (see Recipes)
4. TMRE working solution (see Recipes)
5. MitoTrackerTM Green FM working solution (see Recipes)
Recipes
1. Complete RPMI 1640 cell culture medium
| Reagent | Final concentration | Volume |
|---|---|---|
| RPMI 1640 medium | — | 445 mL |
| FBS | 10% | 50 mL |
| Penicillin-streptomycin | 1× | 5 mL |
Prepare under sterile conditions and store at 4 °C.
2. Phenol red-free imaging medium
| Reagent | Final concentration | Volume |
|---|---|---|
| RPMI 1640 with L-glutamine, without phenol red | — | 445 mL |
| FBS | 10% | 50 mL |
| Penicillin-streptomycin | 1× | 5 mL |
Prepare under sterile conditions and store at 4 °C.
3. FCCP working solution
Prepare a 10 mM FCCP stock solution by dissolving 1.0 mg of FCCP in 393.4 μL of DMSO and store aliquots at -80 °C. On the day of use, prepare a 100 μM FCCP working solution by diluting the 10 mM stock solution 1:100 in phenol red-free imaging medium. Protect FCCP from light.
4. TMRE working solution
Prepare a 1 mM TMRE stock solution by dissolving 1.0 mg of TMRE in 1.94 mL of DMSO and store aliquots at -80 °C. On the day of use, prepare a 10 μM TMRE working solution by diluting the 1 mM stock solution 1:100 in DMSO. TMRE stock and dilutions must be protected from light.
5. MitoTrackerTM Green (MTG) working solution
Prepare a 1 mM MTG stock solution by dissolving 50 μg of lyophilized MTG in 74.42 μL of DMSO and store aliquots at -80 °C. On the day of use, prepare a 10 μM MTG working solution by diluting the 1 mM stock solution 1:100 in DMSO. MTG stock and dilutions must be protected from light.
Laboratory supplies
1. Glass-bottom dishes, 35 mm, #1.5 glass (170 μm +20/-10 μm), D 263 M Schott glass, sterile (Ibidi, catalog number: 81218)
2. 100 × 20 mm tissue culture EasyDishTM with vent (Thermo Fisher Scientific, catalog number: 150466)
Equipment
1. Zeiss Axio Observer Z1 microscope (Carl Zeiss, Jena, Germany) equipped with a Yokogawa CSU-X1 spinning-disk confocal unit (Yokogawa, Tokyo, Japan), a Zeiss Axiocam 506 mono camera, and a Zeiss Plan-Apochromat 63×/1.4 Oil DIC M27 objective. Excitation was provided by 488 and 561 nm laser lines. Fluorescence emission was collected using a Semrock BrightLine quad-band bandpass filter (FF01-446/523/600/677-25; Semrock, Rochester, NY, USA), with four transmission bands: 429.5–462.0 nm (band 1), 502.5–544.5 nm (band 2), 582.0–617.5 nm (band 3), and 663.0–690.5 nm (band 4). Microscope acquisition was controlled using Zeiss ZEN 2 Blue Edition software. Temperature, CO2, and humidity were controlled using PeCon incubation components (PeCon, Erbach, Germany). A humidified cell culture incubator was maintained at 37 °C, 5% CO2, and 90% humidity (Mandel Scientific, model: Heracell)
2. Thermostatic water bath maintained at 37 °C (Fisher Scientific, model: Isotemp GPD 10, catalog number: FSGPD10)
3. Class II Biosafety Cabinet (The Baker Company, model: SterilGARD® III Advance, SG403A)
Software and datasets
1. ZEN 2 Blue Edition microscopy software (Carl Zeiss Microscopy, version 3.12; Jena, Germany)
2. Fiji (ImageJ version 1.54p; National Institutes of Health, USA)
3. Microsoft Excel (Microsoft Corporation, version 2611, Redmond, WA, USA)
4. GraphPad Prism (GraphPad Software, Inc., version 11, Boston, MA, USA)
5. All data and code have been deposited at Codeberg.org: https://codeberg.org/VincentPoupart/ZoeAnneSophie2026
Procedure
A. Experiment 1: Validation of TMRE and MitoTrackerTM Green (MTG) signals and microscopy parameter optimization
Note: This first experiment establishes the cell seeding, staining, and microscopy acquisition conditions required for subsequent quantitative measurements.
1. Cell seeding optimization
a. On day 1, establish the appropriate seeding density to achieve 40%–60% confluence and 2–8 cells per 63× image on the day of imaging. For this protocol, MeWo melanoma cells were seeded at 25,000 cells/mL (or 50,000 cells per 35-mm dish) and cultured for 72 h before imaging. Sufficient time should be allowed after trypsinization for cells to adhere and spread, facilitating visualization and imaging of the mitochondrial network. In preliminary experiments, we recommend testing culture periods of 24–72 h to determine optimal seeding density for the cell line of interest.
b. Once the optimal seeding density has been determined, seed cells in 2 mL of complete growth medium in 35-mm glass-bottom dishes 24–72 h before imaging. Ensure that the dishes are compatible with the confocal microscope, such as the Zeiss Axio Observer Z1 spinning disk system employed in this protocol.
Critical: Avoid overconfluent cultures, as excessive cell density may alter mitochondrial network organization and membrane potential, promote cell overlap, and complicate image acquisition and analysis.
c. Place dishes in an incubator at 37 °C, 5% CO2, and ≥90% relative humidity.
2. Image acquisition: Microscope preparation and environmental stabilization
Note: The objective here is to determine TMRE and MTG imaging conditions that provide a strong fluorescence signal without saturation, excessive background, or visible phototoxicity. Once optimized, save all acquisition settings in the microscope acquisition software and keep them constant throughout the experiment. In this protocol, acquisition settings were saved and applied using ZEN 2 Blue Edition (Zeiss).
a. Before bringing the cells to the microscope room, turn on the confocal microscope, camera, illumination system, and environmental chamber to allow sufficient time (1 h) for the system to stabilize.
b. Set the environmental chamber to 37 °C and 5% CO2.
Critical: Using a test slide or non-critical sample, perform preliminary focusing with the 63× oil-immersion objective, adjust the field diaphragm, and verify that the imaging system is functioning properly. This warm-up period is important because live-cell imaging is highly sensitive to temperature fluctuations, focus drift, and unstable CO2 levels.
Critical: Once the environmental chamber has stabilized at 37 °C and 5% CO2, transfer the cells from the tissue culture incubator to the microscope chamber and allow them to equilibrate for approximately 10 min before staining. This equilibration period minimizes changes in temperature and medium pH and allows imaging conditions to stabilize.
3. Image acquisition: Preparation of imaging medium and fluorescent probes
Critical: For all live-cell imaging experiments, use phenol red-free medium supplemented with serum. Phenol red should be avoided as it may increase background fluorescence and interfere with quantitative fluorescence measurements.
a. Prepare the TMRE and MTG working solutions (see Recipes) and the TMRE+MTG staining medium by diluting the working solutions in phenol red-free imaging medium (Table 1).
Table 1. Preparation of the TMRE+MTG staining medium
| Component | Volume (μL) | Final concentration (nM) |
|---|---|---|
| Phenol red-free imaging medium (Recipe 2) | 1970 | |
| TMRE 10 μM working solution (Recipe 4) | 20 | 100 |
| MTG 10 μM working solution (Recipe 5) | 10 | 50 |
| Final volume | 2,000 |
Critical: These final concentrations should be considered as a starting point and may require optimization depending on the cell line and microscope sensitivity. In our case, the indicated concentrations were optimized for the MeWo melanoma cell line.
Critical: Always use a freshly thawed aliquot for each experiment, as repeated freeze/thaw cycles may compromise the stability and performance of TMRE and MTG.
4. Image acquisition: Live-cell staining procedure and initial microscope setup
a. Once the microscope chamber is stable and the cells have re-equilibrated, remove the culture medium and replace it with 2 mL of prewarmed TMRE+MTG staining medium.
Critical: Preheat the medium to a temperature of 37 °C before use.
b. Return the dish to the environmental chamber and incubate for 30 min at 37 °C and 5% CO2.
c. During this incubation period, finalize the microscope setup in the imaging system and prepare the acquisition protocol (Table 2). Avoid unnecessary fluorescence exposure to minimize photobleaching and phototoxicity before image acquisition.
d. Following the 30-min incubation, wash the cells twice with phenol red-free imaging medium (Recipe 2) and replace with fresh phenol red-free imaging medium before imaging. At this stage, cells are ready for image acquisition.
Table 2. Light intensity, exposure time, and gain used for each channel
| Channel | Laser wavelength (nm) | Light intensity (%) | Exposure time (ms) | EM gain | Number of z-slices |
|---|---|---|---|---|---|
| Cy3 (TMRE) | 555 | 2 | 50 | 200 | 16 |
| EGFP (MTG) | 488 | 40 | 200 | 150 | 16 |
| Brightfield | - | 25 | 50 | 106 | 16 |
Critical: These settings should be considered a starting point only. The objective is to obtain sufficient TMRE and MTG fluorescent signals while avoiding saturation, excessive background, photobleaching, and phototoxicity. Adjust exposure time and lamp intensity before modifying dye concentrations.
5. Image acquisition: Optimization
Note: The objective of this step is not to generate biological data but to establish robust and reproducible acquisition parameters.
a. Use the brightfield channel to identify morphologically healthy, well-separated cells and select fields containing approximately 2–8 cells. Avoid dead, detached, overlapping, or highly confluent cells. Fluorescence channels should not be used for routine sample navigation or focusing before image acquisition, to minimize photobleaching and phototoxicity.
b. Focus using the 63× oil objective and acquire images as z-stacks, with 16 z-slices per field. This parameter may be adjusted according to cell thickness but should remain constant throughout the experiment.
c. Optimize acquisition settings to obtain a robust TMRE signal in control cells. If signals are too high, reduce exposure time or illumination intensity first. If the signal remains too high, reduce TMRE or MTG concentration. If signals are too weak, increase exposure time or lamp intensity carefully. If the signal is still weak, increase TMRE or MTG concentrations cautiously.
Critical: For the initial optimization, prepare 3–5 dishes to test different TMRE and MTG concentrations to identify optimal staining conditions.
Note: The thickness of the z-stack is specific to each cell. It is not necessary to acquire an excessive number of z-slices, as doing so is not relevant to the experiment; 16 z-slices are sufficient, using a predefined z-step.
d. Once optimal staining and acquisition settings have been established, save the protocol in the imaging system and keep all parameters constant for subsequent control and FCCP-treated samples.
B. Experiment 2: Validation of TMRE-based mitochondrial membrane potential measurements using FCCP
Note: After establishing the optimal staining and acquisition conditions in Experiment 1, this second experiment validates the assay using FCCP-induced mitochondrial depolarization and defines the conditions used for quantitative analysis.
1. Cell seeding:
a. After establishing the optimal seeding density in Experiment 1 (approximately 40%–60% confluence with multiple isolated colonies of 2–8 cells), prepare five glass-bottom 35-mm dishes for FCCP depolarization optimization.
b. Seed cells in 2 mL of complete growth medium in 35-mm glass-bottom dishes 24–72 h before imaging.
c. Place dishes in an incubator at 37 °C, 5% CO2, and ≥90% relative humidity.
2. Image acquisition: FCCP depolarization optimization
Objective: FCCP is used as a positive control to depolarize the mitochondrial membrane, thereby confirming that the TMRE signal depends on membrane potential. FCCP concentration and incubation time may require optimization depending on the cell line. The goal is to induce a clear reduction in TMRE fluorescence while preserving sufficient cell viability and mitochondrial content for imaging and quantification. However, excessive loss of mitochondrial signal, widespread cell detachment, or obvious signs of cell death may indicate overly harsh treatment conditions and complicate interpretation, as reduced TMRE fluorescence may reflect loss of cell viability or mitochondrial integrity rather than a specific change in mitochondrial membrane potential. As a guideline, aim for a 60%–80% reduction in TMRE fluorescence relative to untreated cells while maintaining detectable MTG staining and acceptable cell viability.
a. Before bringing the cells to the microscope room, turn on the confocal microscope, camera, illumination system, and environmental chamber to allow sufficient time (1 h) for the system to stabilize.
b. Set the environmental chamber to 37 °C and 5% CO2.
Critical: Using a test slide or non-critical sample, perform preliminary focusing with the 63× oil-immersion objective, adjust the field diaphragm, and verify that the imaging system is functioning properly. This warm-up period is important because live-cell imaging is highly sensitive to temperature fluctuations, focus drift, and unstable CO2 levels.
Critical: Once the environmental chamber has stabilized at 37 °C and 5% CO2, transfer the cells from the tissue culture incubator to the microscope chamber and allow them to equilibrate for approximately 10 min before staining. This equilibration period minimizes changes in temperature and medium pH and allows imaging conditions to stabilize.
3. Image acquisition: Microscope preparation and environmental chamber stabilization
Note: As described in Experiment 1, prepare the microscope, environmental chamber, and imaging workstation before bringing the cells from the incubator.
a. Use the TMRE and MTG concentrations optimized during Experiment 1. Prepare the control staining solution exactly as established during Experiment 1.
b. Prepare different working solutions of FCCP: 1, 2.5, 5, and 10 μM dilutions in phenol red-free imaging medium in 2 mL tubes (Table 3).
c. Incubate cells with the different concentrations of FCCP for 30 min at 37 °C and 5% CO2.
d. Add TMRE and MTG to reach the final concentrations established during Experiment 1 (Table 3).
e. Add dyes and gently agitate the dish in a cross-shaped motion to ensure homogeneous distribution.
f. Incubate for an additional 30 min at 37 °C and 5% CO2.
g. Wash cells twice with phenol red-free imaging medium supplemented with serum.
h. Replace with phenol red-free imaging medium and proceed immediately to image acquisition using the same settings established during Experiment 1.
Table 3. Preparation of FCCP+TMRE+MTG staining solution
| Reagent | Working solution concentration (μM) | Final concentration (μM) | Volume of reagent (μL) | Volume of phenol red-free imaging medium (μL) | Incubation time (min) |
| FCCP | 100 | 1 | 20 | 1,950 | First: 30 |
| 2.5 | 50 | 1,920 | |||
| 5 | 100 | 1,870 | |||
| 10 | 200 | 1,770 | |||
| TMRE | 10 | 0.1 | 20 | Second: 30 | |
| MTG | 10 | 0.05 | 10 |
Note: Because TMRE and MTG are added after FCCP pre-treatment, the FCCP concentration during the initial incubation is 1.5% higher than the final concentration after adjustment to 2 mL. This small difference is acceptable, provided that the workflow is applied consistently across all FCCP-treated samples.
4. Image acquisition: Acquire all images using the exact microscope settings established during Experiment 1. Exposure time, illumination intensity, gain, z-stack parameters, and all other acquisition settings must remain identical between untreated and FCCP-treated samples. For each condition, acquire approximately 15–25 fields of view containing 2–8 cells per field, ensuring that individual cells can be clearly distinguished.
5. FCCP optimization and expected outcome
a. Following FCCP treatment, TMRE fluorescence should be visibly reduced. MTG signal should remain detectable and continue to allow mitochondrial visualization and cell identification.
b. Compare FCCP-treated cells with untreated controls acquired using identical settings. A successful FCCP condition should reduce the TMRE fluorescence while preserving acceptable cell morphology.
Note: If TMRE fluorescence does not decrease sufficiently, increase FCCP concentration and/or incubation time before adding TMRE/MTG. Conversely, if cells become rounded, detached, or show major loss of MTG signal, the FCCP treatment may be too harsh and should be reduced.
c. Once an optimal FCCP condition has been identified, keep FCCP concentration, incubation time, staining conditions, and acquisition settings constant for all future experiments.
6. Final experimental design after optimization: Once seeding density, staining concentration, imaging settings, and FCCP conditions are optimized, the final experiment should be repeated at least three times independently.
For each independent experiment, acquire approximately 15–25 fields of view per condition (Table 4). Each image should ideally contain 2–8 cells. Avoid selecting only the best-looking cells. The fields should be representative, but exclude technically unusable areas, such as out-of-focus regions, dead cells, large debris, saturated fields, or overly confluent clusters.
Table 4. Different conditions used for the experiment
| Condition | Purpose |
|---|---|
| Control (DMSO)/no FCCP | Baseline mitochondrial membrane potential |
| FCCP-treated | Depolarized control |
C. Fiji image processing and quantification
Overview: Each ZEN image contains two channels and a z-stack (approximately 16 z-slices) (Table 2). The following ImageJ scripts were developed to analyze .czi files generated using the ZEN 2 (Blue Edition) microscopy acquisition software and are recommended to ensure standardized and reproducible image processing and analysis.
1. Split channels and sum slices Z-projection
a. All images produced have approximately 16 z-slices and two channels.
b. Split the channels: This will generate a separate image for each channel. You should obtain an EGFP channel corresponding to MTG (C1) and a Cy3 channel corresponding to TMRE (C2).
c. To enable fluorescence quantification, each TMRE or MTG z-stack must be converted into a single 2D image.
d. The function Sum Slices Z-projection reduces the 3D z-stack (approximately 16 z-slices) to a single 2D image by summing the pixel intensity values at each x-y position across all slices along the z-axis. Next, this single image can be quantified with the script (Figure 1A).

Figure 1. Fiji-based image segmentation and fluorescence quantification workflow for MTG and TMRE analysis. (A) Representative MTG (C1) and TMRE (C2) images of control MeWo cells following import of the .czi file into Fiji, sum-slice z-projection, and channel separation. The EGFP (MTG) channel was used to define regions of interest (ROIs) independently of TMRE fluorescence intensity, whereas the corresponding Cy3 (TMRE) channel was subsequently used for fluorescence quantification within the MTG-derived ROIs. (B) A duplicate of the MTG image was processed using a Gaussian blur (σ = 1 μm), automatically thresholded using the Li method, and converted into a binary mask. The red overlay illustrates the pixels selected by Li thresholding, and the resulting binary mask identifies the segmented MTG-positive regions. (C) ROIs were generated from the MTG binary mask using Fiji’s Analyze Particles function with a minimum particle size of 100 μm2. The resulting ROIs were stored in the ROI Manager and overlaid onto the original MTG image to verify their spatial correspondence with MTG-positive regions. (D) The same ROIs generated from the EGFP channel were applied to the corresponding MTG and TMRE images. Mean fluorescence intensity was measured separately in each channel within identical ROIs, allowing TMRE fluorescence to be quantified within regions defined independently from the TMRE signal. (E) Background regions were identified independently for the EGFP and Cy3 channels using automatic MinError thresholding. The resulting background ROIs were applied to their corresponding fluorescence channels to determine the mean background fluorescence intensity. Background-corrected fluorescence was calculated separately for each channel as: corrected mean fluorescence intensity = mean ROI fluorescence intensity - mean background fluorescence intensity. The background-corrected TMRE and MTG measurements were subsequently used to calculate the TMRE/MTG fluorescence ratio.
2. MitoTracker Green segmentation
a. A Gaussian filter is applied to a duplicate of the MTG image to reduce image noise and smooth local intensity variations, thereby enhancing the accuracy of mitochondrial segmentation. The filtered image is successively analyzed using the Li automatic thresholding algorithm and converted into a binary mask (Figure 1B). In this binary mask, white pixels correspond to mitochondrial structures identified based on MTG fluorescence, whereas black pixels represent the background (Figure 1B). The resulting binary mask serves as the basis for describing mitochondrial ROIs, which are applied to the corresponding MTG and TMRE channels for fluorescence intensity quantification (Figure 1B).
b. The binary mask is used to generate ROIs equivalent to MTG-positive mitochondrial structures. Using Fiji’s Analyze Particles function, objects with an area greater than 100 μm2 are selected and converted into ROIs. These ROIs are then saved and applied to the respective MTG and TMRE images to measure fluorescence intensities precisely within mitochondrial regions (Figure 1C, D).
3. Quantification of mitochondrial membrane potential normalized to mitochondrial compartment fluorescence following FCCP-induced depolarization: MTG fluorescence provides a mitochondrial reference signal within the same ROIs. For each cell, mean TMRE fluorescence intensity is quantified within MTG-derived mitochondrial ROIs and expressed relative to the corresponding MTG fluorescence intensity. The ratio helps to account for differences in mitochondrial labeling and mitochondrial area between cells. TMRE and MTG fluorescence values should also be retained and analyzed separately, as experimental treatments may independently affect either signal. Data are quantified on a cell-by-cell basis for each acquired image (Figure 1E).
Data analysis
Representative fluorescence images were acquired from control cells and cells treated with the mitochondrial uncoupler FCCP to illustrate FCCP-induced changes in mitochondrial membrane potential (Figure 2). FCCP treatment resulted in a marked reduction in TMRE fluorescence, whereas MTG fluorescence did not show a statistically significant change. TMRE and MTG fluorescence signals were quantified independently in individual cells using Fiji software, according to the image analysis workflow described above (Figure 1). Mean fluorescence intensity was measured within each cell-defined region of interest, and background fluorescence was subtracted from the signal.
Because experimental treatments may affect MTG fluorescence independently of mitochondrial content, background-corrected TMRE and MTG fluorescence intensities were initially analyzed separately to assess the effect of FCCP treatment on each signal (Figure 3A–D). Independent analysis of the biological replicate means indicates a progressive decrease in TMRE fluorescence with increasing FCCP concentrations (Figure 3B), whereas no significant change in MTG fluorescence was shown across the FCCP concentrations tested (Figure 3D). For statistical analysis, cell-level fluorescence measurements were averaged within each condition for each of the three independent biological experiments. These experiment-level means were used as biological replicates and compared using one-way ANOVA followed by Dunnett’s multiple-comparisons test. For each analyzed cell, the background-corrected TMRE fluorescence intensity was divided by the corresponding background-corrected MTG fluorescence intensity to calculate the TMRE/MTG ratio, which was used as a relative measure of mitochondrial membrane potential. Distributions of individual-cell TMRE/MTG ratios, pooled from the three independent experiments, are presented in Figure 3E for visualization purposes only; individual cells were not considered independent biological replicates for inferential statistical analyses. Ratios were normalized to control within each experiment, and biological replicate means are presented in Figure 3F.

Figure 2. Two-channel imaging of TMRE and MTG fluorescence in control and FCCP-treated cells. Representative fluorescence images of control MeWo cells and cells treated with 5 μM FCCP, showing MTG, TMRE, and merged channels. MTG fluorescence was used to identify the mitochondrial compartment, whereas TMRE fluorescence reflects mitochondrial membrane potential (ΔΨm). A reduction in the TMRE fluorescence in cells treated with 5 μM FCCP signifies mitochondrial depolarization. Images were acquired using a 63× objective as 16-plane z-slices under identical environmental conditions and acquisition settings, including exposure time, illumination intensity, and detector gain, within each independent experiment. The images shown are representative of biological replicates from three independent experiments performed on separate days and using different cell passages. Subsequent quantitative fluorescence analyses were performed using unmodified source images, as described in Figure 1. Scale bars, 20 μm.

Figure 3. Quantification of mitochondrial membrane potential, normalized to mitochondrial compartment fluorescence following FCCP-induced depolarization. Mitochondrial membrane potential (ΔΨm) was quantified using the TMRE/MTG ratio. Mitochondrial regions of interest (ROIs) were identified from the EGFP (MTG) fluorescence channel (C1), independently of the Cy3 (TMRE) fluorescence intensity (C2) to avoid bias in ROI selection. Background-corrected MTG and TMRE fluorescence intensities were calculated for each analyzed cell as the mean ROI fluorescence measured within the corresponding field. TMRE/MTG ratios were calculated for each cell. (A) Background-corrected TMRE fluorescence intensities of individual cells following treatment with increasing FCCP concentrations. (B) Mean background-corrected TMRE fluorescence intensity for each independent biological experiment. (C) Background-corrected MTG fluorescence intensities of individual cells following treatment with increasing FCCP concentrations. (D) Mean background-corrected MTG fluorescence intensity for each independent biological experiment. (E) TMRE/MTG fluorescence ratios of individual cells following treatment with increasing FCCP concentrations. (F) TMRE/MTG ratios were normalized to the mean control (Ctrl) value within each independent biological experiment and expressed as fold change relative to the control condition. (A, C, E) Individual cell measurements from three independent biological experiments are shown for visualization. Each point represents one analyzed cell. 100–250 cells were analyzed. Data are presented as mean ± SD. (B, D, F) Each point represents one independent biological experiment (n = 3), and bars represent mean ± SEM. Statistical significance was assessed with one-way ANOVA followed by Dunnett’s multiple-comparison test, with experimental conditions matched within each independent biological experiment. *p < 0.05; ns: not significant. Microsoft Excel was used to calculate the mean values presented in B and D, and the TMRE/MTG fluorescence intensity ratio. GraphPad Prism was used to calculate fold changes, generate the graphs, and perform statistical analyses.
Validation of protocol
This protocol was validated in three independent experiments performed on separate days. In each experiment, control melanoma cells and cells treated with 1, 2.5, or 5 μM FCCP were stained for 30 min at 37 °C and 5% CO2 with TMRE and MTG and imaged using identical confocal microscopy acquisition parameters within each independent experiment. For each condition, 15–25 randomly selected non-overlapping fields were acquired. Fields were included in the analysis when they met the image quality and cell density criteria, resulting in approximately 100–250 individual cells analyzed per condition in three independent experiments taken together.
Across all biological replicates, FCCP treatment induced a dose-dependent decrease in TMRE fluorescence and in the TMRE/MTG fluorescence ratio, consistent with mitochondrial membrane depolarization (Figure 3B, F). In contrast, MTG fluorescence did not differ significantly across the FCCP concentrations tested (Figure 3D). Similar reductions were observed in each independent experiment, supporting the reproducibility of the workflow.
To minimize analytical variability, we processed all images using identical segmentation, background correction, ROI generation, and fluorescence quantification parameters. The automated Fiji-based analysis workflow applied identical processing and quantification parameters to all images, thereby limiting user-dependent variation during data extraction and image analysis between experimental groups. We assessed statistical significance using one-way ANOVA followed by Dunnett’s multiple-comparison test.
The protocol is an optimized version of the protocol used and validated in the following research article:
• Lefrançois et al. [24]. The role of ATP synthase subunit e (ATP5I) in mediating the metabolic and antiproliferative effects of metformin in cancer cells. eLife. (Figure 6D, E).
General notes and troubleshooting
General notes
1. On the day of imaging, replace the culture medium with a prewarmed imaging medium compatible with live-cell fluorescence microscopy. A phenol red-free medium is recommended to reduce background fluorescence and improve signal clarity, enhancing the quantitative measurements. Cells should also be cultured and imaged in a microscopy-compatible dish equipped with an optical glass bottom matched to the objective and imaging system used, ensuring optimal results during confocal microscopy.
2. Cell density influences the reliability of fluorescence measurements. The cell density established during optimization should be kept consistent across all experiments to minimize variability. Cultures should be maintained at approximately 40%–60% confluence, and multiple non-overlapping fields should be acquired using a 63× objective. Also, imaging fields should contain 2–8 isolated single cells with clearly distinguishable cell boundaries. Fields with densely packed cell clusters or overlapping cells should be avoided for imaging. Individual cells must be distinguished, and mitochondrial structure must be properly segmented. Small clusters may be included if cell boundaries can be clearly identified and each cell can be analyzed independently with Fiji’s macro.
3. Identical microscope acquisition settings should be maintained across all biological samples. After optimization, the acquisition sequence, detector gain, exposure time, illumination intensity, objective, number of z-slices, and z-step should be kept constant across all experimental conditions within each independent experiment. Then, the same acquisition settings should be preserved across biological replicates to enhance reproducibility. Since fluorescence measurements are normalized to the corresponding control within each independent experiment, quantitative comparisons should be done using normalized rather than raw fluorescence intensity values across replicates.
4. Detector saturation should be avoided as an acquisition criterion. Saturated pixels do not provide reliable quantitative information on fluorescence intensity. To avoid saturation, exposure time and illumination intensity should be adjusted to ensure that the fluorescence signal remains within the camera’s dynamic range across all experimental conditions, including conditions that may increase TMRE fluorescence relative to controls.
5. Fluorescence exposure should be minimized before image capture. For sequential fluorescence acquisition, the TMRE channel should be imaged before the EGFP (MTG) channel to reduce the potential influence of prior excitation on mitochondrial membrane potential measurements. This order of acquisition should be maintained consistently across all experimental conditions. Because repeated fluorescence exposure can promote photobleaching and phototoxicity, sample navigation and focusing should be done using the brightfield channel.
6. Stable environmental conditions should be maintained throughout live-cell imaging. Temperature, CO2, and pH fluctuations can alter cellular and mitochondrial physiology. The environmental chamber should be pre-equilibrated at 37 °C and 5% CO2 for 1 h before acquisition to ensure optimal system stability. Predefined stage positions and the autofocus routine should be run before image acquisition to allow stabilization of the microscope and environmental control system.
7. FCCP, MTG, and TMRE should be prepared as single-use aliquots to minimize freeze/thaw cycles and contamination of the stock solutions. MTG and TMRE are light-sensitive and should be stored and handled protected from light, using amber tubes or tubes wrapped in aluminum foil. When possible, a freshly thawed aliquot should be used for each independent experiment.
8. Randomize or standardize field selection. Fields should be selected according to predefined technical criteria, independently of fluorescence intensity or mitochondrial morphology. Approximately 15–25 fields should be acquired per condition.
9. MTG is commonly used as a membrane potential–independent marker of mitochondrial content, but its stability should be verified under the experimental conditions. Changes in MTG fluorescence following membrane potential–altering treatments, including FCCP, have occasionally been reported and may depend on the cell type and experimental conditions [17,20,28]. MTG fluorescence should therefore be quantified and reported independently, and its stability under the experimental conditions should be verified before interpreting TMRE/MTG ratios. If MTG fluorescence is substantially affected by the experimental treatment, alternative approaches for assessing mitochondrial content should be considered. Alternative membrane potential–independent markers of mitochondrial content include mitochondrial proteins such as TOM20, VDAC1, or citrate synthase. However, these approaches generally require fixation or analysis of separate samples and therefore cannot replace MTG for the simultaneous assessment of mitochondrial content and membrane potential in individual live cells described in this protocol. They may nevertheless provide useful complementary validation when experimental conditions are suspected to affect MTG fluorescence independently of mitochondrial content.
Troubleshooting
Problem 1: TMRE fluorescence intensity is interpreted as an absolute measurement of mitochondrial membrane potential (ΔΨm).
Possible causes: Factors such as dye concentration and loading, mitochondrial content, fluorescence attenuation, imaging parameters, and experimental conditions can affect TMRE fluorescence. Therefore, TMRE fluorescence intensity cannot be directly converted into absolute ΔΨm values (expressed in millivolts) without appropriate calibration.
Solutions: Analyze TMRE fluorescence as a relative indicator of mitochondrial membrane polarization. Perform comparisons only between samples analyzed under identical conditions and include FCCP-treated cells as a depolarization control. Report TMRE fluorescence as relative or normalized fluorescence intensity, reflecting differences in mitochondrial membrane polarization between experimental conditions, rather than as absolute ΔΨm values [14,22,23].
Problem 2: TMRE fluorescence intensity is weak in control cells.
Possible causes: Insufficient TMRE loading, excessive removal of intracellular dye during washing, loss of dye integrity due to light exposure or improper storage, or reduced mitochondrial polarization associated with decreased cell viability may cause low TMRE fluorescence intensity.
Solutions: Verify cell viability, TMRE preparation, and storage conditions. Optimize excitation intensity or exposure settings if fluorescence remains low before increasing TMRE concentration. Maintain identical imaging and staining conditions once an adequate signal is obtained.
Problem 3: TMRE fluorescence is saturated.
Possible cause: TMRE concentration is excessive or imaging parameters are set too high (e.g., exposure time, detector gain, or illumination intensity), resulting in fluorescence saturation and reduced sensitivity.
Solutions: Optimize acquisition settings by reducing exposure time, detector gain, and illumination intensity. Decrease TMRE concentrations only if intensity saturation persists. Exclude saturated images from image quantification analyses. Fluorescence intensities must be established within the detector’s linear dynamic range of the microscope.
Problem 4: FCCP treatment fails to produce the expected loss of TMRE fluorescence associated with mitochondrial depolarization.
Possible causes: FCCP has lost activity because of degradation or inadequate storage; TMRE concentration is suboptimal; or treatment conditions are insufficient to induce complete mitochondrial depolarization. Acquisition settings can also prevent detection of FCCP-induced alterations in TMRE signal.
Solutions: Prepare a fresh FCCP working solution from a properly stored stock solution. Verify that the FCCP concentration and incubation time are sufficient to induce mitochondrial depolarization in the cell type of interest. Ensure that fluorescence signals in control cells are not saturated and remain within the detector’s linear dynamic range. Because FCCP sensitivity varies across cell types and experimental conditions, both FCCP conditions and exposure time should be optimized for each experimental model. An initial dose-response titration is recommended to identify the lowest FCCP concentration that induces substantial mitochondrial depolarization within the desired experimental timeframe. While some cancer cell lines may require higher FCCP concentrations to achieve complete mitochondrial depolarization, more sensitive cell types should be evaluated using lower FCCP concentrations. Excessive FCCP concentrations or prolonged incubation periods should be avoided, as they can cause secondary cellular effects, compromise cell viability, and confound the interpretation of mitochondrial membrane potential analysis [29–31].
Problem 5: Photobleaching occurs during image acquisition.
Possible causes: High illumination intensity, repeated focusing under fluorescence illumination, prolonged exposure, or repeated acquisition of the same field are associated with excessive light exposure and progressive photobleaching.
Solutions: Minimize illumination intensity and exposure time while preserving an adequate fluorescence signal for quantitative analysis. Perform focusing and navigation using the brightfield (transmitted light) channel and limit fluorescence exposure before image acquisition. Avoid repeated imaging of the same field and keep acquisition settings identical across all replicates.
Problem 6: Image-based segmentation of mitochondrial structures using thresholding is inaccurate.
Possible causes: Insufficient contrast between mitochondrial signal and background, inappropriate thresholding parameters, heterogeneous background fluorescence, or excessive segmentation of interconnected mitochondrial structures during image processing.
Solutions: Optimize staining and image acquisition before modifying segmentation parameters. Validate background subtraction and thresholding using representative images, then apply the same analysis workflow to all replicates. If global thresholding does not reliably distinguish mitochondrial structures from the background, use an alternative segmentation approach that has been verified by visual inspection to clearly identify mitochondrial architecture while minimizing background inclusion, artificial fragmentation, and merging artifacts. Acquisition settings must stay consistent between experimental conditions and biological replicates.
Acknowledgments
This work was supported by the Natural Sciences and Engineering Research Council of Canada (Grant RGPIN-2023-03828, 2023). We thank Christian Charbonneau (Bio-Imaging Core Facility, IRIC, University of Montreal) for technical assistance. This protocol is adapted and modified from the article: Lefrançois, Guillaume et al. “The role of ATP synthase subunit e (ATP5I) in mediating the metabolic and antiproliferative effects of metformin in cancer cells.” eLife vol. 13 RP102680. 15 May. 2026, doi:10.7554/eLife.102680 [24].
Graphical overview was created in BioRender: Marceau, A. (https://BioRender.com/o6b763q), licensed under CC BY 4.0.
Author contributions
Anne-Sophie Marceau and Zoé Lidy developed the protocol, performed experiments, analyzed the results, and wrote the manuscript; Guillaume Lefrançois contributed to protocol development; Vincent Poupart and Eugénie Goupil developed the algorithm; Gerardo Ferbeyre and Andreea R. Schmitzer supervised Guillaume Lefrançois and edited the manuscript; Simon-Pierre Gravel conceptualized the protocol, supervised protocol development and experimental validation, designed, wrote, and edited the manuscript, and secured funding.
Competing interests
The authors declare no conflicts of interest.
Ethical considerations
This study did not involve human subjects or live animals. Therefore, ethics committee approval and informed consent were not required.
References
文章信息
稿件历史记录
提交日期: Sep 18, 2026
接收日期: Nov 8, 2026
在线发布日期: Nov 24, 2026
出版日期: Dec 5, 2026
版权信息
© 2026 The Author(s); This is an open access article under the CC BY license (https://creativecommons.org/licenses/by/4.0/).
如何引用
Readers should cite both the Bio-protocol article and the original research article where this protocol was used:
您对这篇实验方案有问题吗?
在此处发布您的问题,我们将邀请本文作者来回答。同时,我们会将您的问题发布到Bio-protocol Exchange,以便寻求社区成员的帮助。
Share
Bluesky
X
Copy link