The determined buffer conductivity and flow rate conditions remained the same for IM-C units since they were integrated with DEP-D unit. IM-C units were utilized for cell counting. The working frequency was determined as 100 kHz to be able to count cells while DEP-D unit was also under operation with voltages at 6.2 MHz and 8.6 MHz frequencies. Voltage magnitude was determined as 2Vpp. With voltage values under 2Vpp, IM-C units did not work due to the interference between IM-C units' excitation signals and the signals used in the DEP-D unit.