2.3.3. Linear range verification
This protocol is extracted from research article:
Performance evaluation of CA242 by flow fluorescence assay
Medicine (Baltimore), Apr 16, 2021; DOI: 10.1097/MD.0000000000025301

Following the EP6-A guideline published by the CLSI,[12] the concentrations of L-Series standard solutions (L1-L5, known concentrations of 15.84 U/ml, 120.34 U/ml, 224.83 U/ml, 329.33 U/ml and 433.82 U/ml, respectively) were detected by the testing kit. All raw fluorescence intensity data were converted into corresponding concentrations through the calibration curve and a scatter diagram was made (X: theoretical concentration, Y: instrumental concentration). Then, multiple regression models were applied to analyze the linear range.