The product analysis was performed on a Dionex Ultimate 3000 Series UPLC (Thermo Scientific, MA, USA) and a 100 × 2.1 mm 1.9 μm Hypersil Gold C18 column (Thermo Scientific, MA, USA) with 100% acetonitrile used as mobile phase A and 0.1% formic acid in ultrapure water used as mobile phase B, running at 0.5 ml/min and 40 °C. The mobile gradient was as follows: 0–2 min, 83% B; 2–5 min, 83–80% B; 5–7 min, 80–75% B; 7–11 min, 75% B; 11–11.1 min, 75–83% B; and 11.1–13 min, 83% B. Detection was performed at 280 nm for eriodictyol and 350 nm for luteolin.