Following irradiation with x-rays, an EPSON Expression 10000XL Graphic Arts scanner was used to scan films before analysis, consistent with clinical practice (44). After scanning, the RIT dosimetry software (RIT113 Classic V6.0) was used to analyze digitized image of the films. Pixel values at discrete points with a 2-mm spacing between each point were measured along the digitized films. Each pixel value is correlated to the predicted dose using clinical standard calibration curve, and a corresponding topographical dose map is acquired. The regions receiving a high radiation dose are depicted in red, and those receiving a low irradiation dose are depicted in blue.